DOI 10.1002/ccd.29700

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2021

Heart valve sizing and clinical outcomes in patients undergoing transcatheter aortic valve implantation.

Okuno Taishi, Heg Dik, Lanz Jonas, Praz Fabien, Gräni Christoph, Langhammer Bettina, Reineke David, Räber Lorenz, Wenaweser Peter, Pilgrim Thomas, Windecker Stephan, Stortecky Stefan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

To investigate the impact of transcatheter heart valve (THV) sizing on procedural results and clinical outcomes following transcatheter aortic valve implantation (TAVI).

BACKGROUND

The impact of individual THV sizing for patients with borderline aortic annulus anatomy remains unclear.

METHODS

In the prospective BernTAVI registry, THV sizing conditions were retrospectively evaluated, and patients were categorized into three groups based on the recommendations and the sizing chart of the manufacturers: optimal sizing, borderline sizing (THV size located within 5% to each border of the optimal sizing recommendation), and suboptimal sizing (THV size outside the recommended range). The latter two groups were further subcategorized into THV-oversizing and THV-undersizing. The primary endpoint was a composite of all-cause death and unplanned repeat intervention at 1 year.

RESULTS

Out of a total of 1,638 patients who underwent TAVI, 9.5 and 15.6% of patients were categorized into the borderline and suboptimal sizing group, respectively. Device success was achieved in 87.4, 88.9, and 83.6% of patients with optimal, borderline, and suboptimal sizing, respectively. The primary endpoint occurred in 12.3% of patients with optimal sizing, 14.9% of patients with borderline sizing (HRadj 1.35, 95%CI 0.87-2.09), and in 17.4% of patients with suboptimal sizing (HRadj 1.42, 95%CI 1.01-1.99). Within the suboptimal sizing cohort, unfavorable outcomes were mainly associated with THV undersizing (device success: 76.4%, primary endpoint: 23.9%, HRadj 1.98, 95%CI 1.36-2.87).

CONCLUSION

Suboptimal TAVI prosthesis sizing is associated with an increased risk of all-cause death and unplanned repeat intervention within 1 year largely attributable to undersized THV prostheses.

Для цитирования: Okuno Taishi, Heg Dik, Lanz Jonas, Praz Fabien, Gräni Christoph, Langhammer Bettina, Reineke David, Räber Lorenz, Wenaweser Peter, Pilgrim Thomas, Windecker Stephan, Stortecky Stefan. Heart valve sizing and clinical outcomes in patients undergoing transcatheter aortic valve implantation.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2021. DOI: 10.1002/ccd.29700.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.