Novel aggregated multiposition noncontact mapping of atrial tachycardia in humans: From computational modeling to clinical validation.

Shi Rui, Zaman Junaid A B, Chen Zhong, Shi Xinwei, Zhu Min, Sathishkumar Anitha, Boyalla Vennela, Karim Nabeela, Cantor Emily, Haldar Shouvik, Jones David G, Hussain Wajid, Markides Vias, Virdee Munmohan, Wang Xunzhang, Grace Andrew, Wong Tom

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

A novel aggregated multiposition noncontact mapping (AMP-NCM) algorithm is proposed to diagnose cardiac arrhythmias.

OBJECTIVE

The purpose of this study was to computationally determine an accuracy threshold and to compare the accuracy and clinical utility of AMP-NCM to gold standard contact mapping.

METHODS

In a cellular automata model, the number of catheter positions and chamber coverage were varied to establish accuracy requirements for clinically relevant AMP-NCM. This guided the clinical study protocol. In a prospective cohort of patients with atrial tachycardia (AT), noncontact mapping (NCM) recordings from a single position (SP) and multiple positions were compared to contact mapping with a high-density multipolar catheter using morphology and timing differences of reconstructed signals. Identification of AT mechanisms and ablation targets using both AMP-NCM and contact mapping were randomly evaluated by 5 blinded reviewers.

RESULTS

AMP-NCM accuracy was asymptotic at 60 catheter positions in computational modeling. Twenty patients (age 65 ± 12 years; 19 male) with 26 ATs (5 focal, 21 reentrant) were studied. Morphologic correlation of signals derived from AMP-NCM was significantly better than those from SP-NCM compared to contact signals (median 0.93 vs 0.76; P <.001). AMP-NCM generated maps more rapidly than contact mapping (3 ± 1 minutes vs 13 ± 6 minutes; P <.001) and correctly diagnosed AT mechanisms in 25 of 26 maps (96%). Overall, 80% of arrhythmia mechanisms were correctly identified using AMP-NCM by blinded reviewers.

CONCLUSION

Once 60 catheter positions were achieved, AMP-NCM successfully diagnosed mechanisms of AT and identified treatment sites equal to gold standard contact mapping in 3 minutes of procedural time.

Для цитирования: Shi Rui, Zaman Junaid A B, Chen Zhong, Shi Xinwei, Zhu Min, Sathishkumar Anitha, Boyalla Vennela, Karim Nabeela, Cantor Emily, Haldar Shouvik, Jones David G, Hussain Wajid, Markides Vias, Virdee Munmohan, Wang Xunzhang, Grace Andrew, Wong Tom. Novel aggregated multiposition noncontact mapping of atrial tachycardia in humans: From computational modeling to clinical validation.. Heart rhythm. 2021. DOI: 10.1016/j.hrthm.2021.09.025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.