Медицинская библиотека · The Journal of heart valve disease · 1998

The Ultracor tilting disc heart valve prosthesis: a seven-year study.

Li H H, Jeffrey R R, Davidson K G, Seifert D, Körfer R, Grunkemeier G L

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND AND AIM OF THE STUDY: The Ultracor heart valve is a recent entry in the evolution of the tilting disc valve. This report summarizes the experience with the Ultracor valve from three European centers.

METHODS

Between 1990 and 1996, 446 patients received 499 Ultracor heart valve prostheses in 450 procedures, including 225 (50.0%) aortic, 172 (38.2%) mitral and 49 (10.9%) double valve replacements. An additional four (1.0%) patients had mitral valve replacement (MVR) added to a previous aortic valve replacement (AVR) and were considered double valve replacement (DVR) patients. The total follow up was 751 patient-years (pt-yr) for AVR (mean 3.4), 440 pt-yr for MVR (mean 2.6) and 125 pt-yr for DVR (mean 2.4). Nine patients (one AVR, eight MVR) were lost to follow up, which was 98% complete.

RESULTS

The actuarial survival rate, including operative mortality rate, at five years was 90% for AVR, 77% for MVR and 82% for DVR. The linearized complication rates (%/year) for AVR, MVR and DVR were: 2.1, 4.0 and 4.0 for late mortality; 0.1, 3.0 and 0.8 for thromboembolism; 0, 0.2 and 0 for thrombosis; 2.0, 1.6 and 1.6 for anticoagulant-related hemorrhage (ACH); 0.3, 0.5 and 1.6 for prosthetic valve endocarditis (PVE); and 0.5, 0.9 and 3.2 for reoperation, respectively. The actuarial rates of freedom from complications at five years were: thromboembolism, 99% for AVR and 88% for MVR; thrombosis, 100% for AVR and 99% for MVR; ACH, 91% for AVR and 94% for MVR; PVE, 99% for AVR and 97% for MVR; reoperation, 98% for AVR and 98% for MVR. No structural failure was observed.

CONCLUSIONS

Seven years' experience showed the Ultracor heart valve prosthesis to be comparable with other currently used mechanical heart valves. Continued evaluation of this prosthesis is warranted in order to obtain a more extended clinical follow up.

Для цитирования: Li H H, Jeffrey R R, Davidson K G, Seifert D, Körfer R, Grunkemeier G L. The Ultracor tilting disc heart valve prosthesis: a seven-year study.. The Journal of heart valve disease. 1998.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.