MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/s0003-4975(00)02652-7

Медицинская библиотека · The Annals of thoracic surgery · 2001

Intermediate-term results with 1,019 carbomedics aortic valves.

Li H H, Hahn J, Urbanski P, Torka M, Grunkemeier G L, Hacker R W

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

A retrospective study was conducted to evaluate the intermediate-term outcome in patients with the Carbomedics aortic valve prosthesis.

METHODS

The study included 1,019 primary valve replacements between 1989 and 1997. Seventy-two percent of patients were men; mean (standard deviation) age was 61 (10) years. The preoperative New York Heart Association functional class was III or IV in 70% of patients. Follow-up at 9 years was 99.6% complete, comprising 2,730 patient-years (mean, 2.7 years).

RESULTS

Patient survival, including operative deaths, was 80% at 7 years. The linearized death rate was 2.9%/year. Statistically significant risk factors for mortality were diabetes, pure valve insufficiency, advanced age at operation, and advanced preoperative functional class. Linearized rates were thrombosis, 0.1%/year; thromboembolism, 1.0%/year; hemorrhage, 1.7%/year; endocarditis, 0.1%/year; paravalvular leak, 0.1%/year; reoperation, 0.1%/year; and all events, 3.0%/year. The 7-year estimates of freedom from complications were thrombosis, 99%; thromboembolism, 93%; hemorrhage, 89%; endocarditis, 99%; paravalvular leak, 99.7%; reoperation, 99%; and all events, 82%. No structural valve failure was observed.

CONCLUSIONS

The low incidence of valve-related complications favors the continued use of the Carbomedics valve in the aortic position.

Для цитирования: Li H H, Hahn J, Urbanski P, Torka M, Grunkemeier G L, Hacker R W. Intermediate-term results with 1,019 carbomedics aortic valves.. The Annals of thoracic surgery. 2001. DOI: 10.1016/s0003-4975(00)02652-7.