MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2019.04.048

Медицинская библиотека · Heart rhythm · 2019

Understanding Outcomes with the EMBLEM S-ICD in Primary Prevention Patients with Low EF Study (UNTOUCHED): Clinical characteristics and perioperative results.

Boersma Lucas V, El-Chami Mikhael F, Bongiorni Maria Grazia, Burke Martin C, Knops Reinoud E, Aasbo Johan D, Lambiase Pier D, Deharo Jean Claude, Russo Andrea M, Dinerman Jay, Shaik Naushad, Barr Craig S, Carter Nathan, Appl Ursula, Brisben Amy J, Stein Kenneth M, Gold Michael R

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The subcutaneous implantable cardioverter-defibrillator (S-ICD) has shown favorable outcomes in large registries with broad inclusion criteria. The cohorts reported had less heart disease and fewer comorbidities than standard ICD populations.

OBJECTIVE

The purpose of this study is to characterize acute performance for primary prevention patients with a left ventricular ejection fraction (LVEF) ≤35% (primary prevention ≤35%).

METHODS

Primary prevention ≤35% patients with no prior documented sustained ventricular tachycardia (VT), pacing indication, end-stage heart failure, or advanced renal failure were prospectively enrolled. Analyses included descriptive statistics, Kaplan-Meier time to event, and multivariable linear and logistic regression.

RESULTS

In 1112 of 1116 patients, an S-ICD was successfully implanted (99.6%). Predictors for longer procedure time included 3-incision technique, higher body mass index (BMI), performing defibrillation testing (DFT), imaging, younger age, black race, and European vs North American centers. Patients undergoing DFT (82%) were successfully converted (99.2%; 93.5% converting at ≤65 J). Higher BMI was predictive of failing DFT at ≤65 J. The rate of 30-day freedom from complications was 95.8%. Most complications involved postoperative healing (45%) or interventions after DFT or impedance check (19%).

CONCLUSION

The procedural outcome data of UNTOUCHED reinforce that S-ICD therapy has low perioperative complication rates and high conversion efficacy of induced ventricular fibrillation, even in a higher-risk cohort with low LVEF and more comorbidities than previous S-ICD studies. Higher BMI warrants more careful attention to implant technique.

Для цитирования: Boersma Lucas V, El-Chami Mikhael F, Bongiorni Maria Grazia, Burke Martin C, Knops Reinoud E, Aasbo Johan D, Lambiase Pier D, Deharo Jean Claude, Russo Andrea M, Dinerman Jay, Shaik Naushad, Barr Craig S, Carter Nathan, Appl Ursula, Brisben Amy J, Stein Kenneth M, Gold Michael R. Understanding Outcomes with the EMBLEM S-ICD in Primary Prevention Patients with Low EF Study (UNTOUCHED): Clinical characteristics and perioperative results.. Heart rhythm. 2019. DOI: 10.1016/j.hrthm.2019.04.048.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.