Simultaneous Comparison of Electrocardiographic Imaging and Epicardial Contact Mapping in Structural Heart Disease.
Аннотация
BACKGROUND
The accuracy of ECG imaging (ECGI) in structural heart disease remains uncertain. This study aimed to provide a detailed comparison of ECGI and contact-mapping system (CARTO) electrograms.
METHODS
Simultaneous epicardial mapping using CARTO (Biosense-Webster, CA) and ECGI (CardioInsight) in 8 patients was performed to compare electrogram morphology, activation time (AT), and repolarization time (RT). Agreement between AT and RT from CARTO and ECGI was assessed using Pearson correlation coefficient, ρ AT and ρ RT, root mean square error, E AT and E RT, and Bland-Altman plots.
RESULTS
After geometric coregistration, 711 (439-905; median, first-third quartiles) ECGI and CARTO points were paired per patient. AT maps showed ρ AT=0.66 (0.53-0.73) and E AT=24 (21-32) ms, RT maps showed ρ RT=0.55 (0.41-0.71) and E RT=51 (38-70) ms. The median correlation coefficient measuring the morphological similarity between the unipolar electrograms was equal to 0.71 (0.65-0.74) for the entire signal, 0.67 (0.59-0.76) for QRS complexes, and 0.57 (0.35-0.76) for T waves. Local activation map correlation, ρ AT, was lower when default filters were used (0.60 (0.30-0.71), P=0.053). Small misalignment of the ECGI and CARTO geometries (below ±4 mm and ±4°) could introduce variations in the median ρ AT up to ±25%. Minimum distance between epicardial pacing sites and the region of earliest activation in ECGI was 13.2 (0.0-28.3) mm from 25 pacing sites with stimulation to QRS interval <40 ms.
CONCLUSIONS
This simultaneous assessment demonstrates that ECGI maps activation and repolarization parameters with moderate accuracy. ECGI and contact electrogram correlation is sensitive to electrode apposition and geometric alignment. Further technological developments may improve spatial resolution.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.