MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ijpharm.2025.125804

Медицинская библиотека · International journal of pharmaceutics · 2025

Subcutaneous depot formation and diffusion in autoinjector delivery: insights from high-speed synchrotron imaging.

Derakhshandeh Rozhin, Meyers Brett A, Zhang Jiacheng, Veilleux Jean-Christophe, Shi Galen H, Vlachos Pavlos P

СтатьяПеревод пока не подготовлен

Аннотация

Autoinjectors are used for the self-administration of subcutaneous (SC) medications, offering controlled, consistent dosing while reducing errors and contamination risk. Despite their advantages in improving patient experience and cost reduction, limited research explores how autoinjector delivery parameters influence drug dispersion and absorption. This study explores the impact of autoinjector delivery parameters on factors including plume growth, spread direction, morphology, and diffusion during injection and within the first minute post-injection. We evaluated three autoinjector models intended to deliver different amounts of product (0.5, 1, and 2 mL) using high-speed synchrotron radiography and computed tomography (CT) on excised pork belly tissue. Synchrotron radiography provided high-resolution, real-time 2D visualization of drug depot formation and diffusion while CT captured 3D images of the post-injection plume. Our analyses reveal that the plume growth rate for all three tested autoinjector models is non-linear, with a rapid initial phase that slows as the injection progresses. On average, the final plume is 25 % larger than the delivered volume due to depot spread within the tissue. For all three models, the plume predominantly expands horizontally (parallel to the tissue), with the aspect ratio increasing throughout the injection and reaching a final value of approximately 4. While the surface area of the plume increases with higher injection volumes, diffusion appears to be minimally influenced by autoinjector and is more likely to be influenced by tissue properties. These insights advance our understanding of autoinjector-tissue interactions, offering critical data to optimize device design and improve computational models for drug absorption and delivery.

Для цитирования: Derakhshandeh Rozhin, Meyers Brett A, Zhang Jiacheng, Veilleux Jean-Christophe, Shi Galen H, Vlachos Pavlos P. Subcutaneous depot formation and diffusion in autoinjector delivery: insights from high-speed synchrotron imaging.. International journal of pharmaceutics. 2025. DOI: 10.1016/j.ijpharm.2025.125804.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.