Enhanced Procedural Efficacy in Typical Atrial Flutter Ablation With a Visualizable Steerable Sheath.

Fuding Guo, Xiaohua Zhao, Sen Yang, Chende He, Mei Lu, Jing Yang, Dongyan Li, Qiwei Liao, Shaolong Li

СтатьяОбзорПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

The Vizigo sheath, a novel visualizable steerable sheath, has been utilized effectively in the clinical management of atrial fibrillation. However, its application in the ablation of typical atrial flutter (AFL) remains unexplored. This study aims to evaluate and compare the efficacy and safety of the Vizigo sheath against a conventional fixed sheath during catheter ablation for typical AFL.

METHODS

This single-center cohort study involved 60 patients undergoing their first AFL ablation procedure. Patients were divided into two groups: the Vizigo sheath group and the fixed sheath group. The primary endpoints included procedural efficiency, lesion quality, and recurrence rates.

RESULTS

The Vizigo sheath demonstrated significant advantages over the fixed sheath, including a shorter procedure time (59.8 ± 13.8 vs. 71.5 ± 11.8 min; p = 0.0008) and reduced fluoroscopy duration (0.96 ± 0.32 vs. 1.31 ± 0.24 min; p < 0.0001). The initial bidirectional block rate was also higher in the Vizigo sheath group (84% vs. 68%). A detailed analysis of the cavotricuspid isthmus (CTI) revealed significant differences in the anterior 2/3 of the CTI, where the Vizigo sheath demonstrated greater catheter contact force, higher ablation index, increased force-time integral, and larger impedance drops (all p < 0.05). In contrast, no significant differences were observed between groups in the posterior 1/3 of the CTI. Both groups achieved a 100% acute procedural success rate, with no major complications reported. During a mean follow-up period of 18.68 months, the AFL recurrence rates were comparable between the two groups (p = 0.75).

CONCLUSIONS

The Vizigo sheath enhances procedural efficiency and lesion quality during AFL ablation, particularly in challenging anatomical regions, and offers notable advantages over traditional fixed sheaths. Further studies are needed to assess its long-term clinical benefits.

Для цитирования: Fuding Guo, Xiaohua Zhao, Sen Yang, Chende He, Mei Lu, Jing Yang, Dongyan Li, Qiwei Liao, Shaolong Li. Enhanced Procedural Efficacy in Typical Atrial Flutter Ablation With a Visualizable Steerable Sheath.. Pacing and clinical electrophysiology : PACE. 2025. DOI: 10.1111/pace.15143.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.