Lymphedema evaluation using noninvasive 3T MR lymphangiography.
Аннотация
PURPOSE: To exploit the long 3.0T relaxation times and low flow velocity of lymphatic fluid to develop a noninvasive 3.0T lymphangiography sequence and evaluate its relevance in patients with lymphedema.
MATERIALS AND
METHODS
A 3.0T turbo-spin-echo (TSE) pulse train with long echo time (TEeffective = 600 msec; shot-duration = 13.2 msec) and TSE-factor (TSE-factor = 90) was developed and signal evolution simulated. The method was evaluated in healthy adults (n = 11) and patients with unilateral breast cancer treatment-related lymphedema (BCRL; n = 25), with a subgroup (n = 5) of BCRL participants scanned before and after manual lymphatic drainage (MLD) therapy. Maximal lymphatic vessel cross-sectional area, signal-to-noise-ratio (SNR), and results from a five-point categorical scoring system were recorded. Nonparametric tests were applied to evaluate study parameter differences between controls and patients, as well as between affected and contralateral sides in patients (significance criteria: two-sided P < 0.05).
RESULTS
Patient volunteers demonstrated larger lymphatic cross-sectional areas in the affected (arm = 12.9 ± 6.3 mm2 ; torso = 17.2 ± 15.6 mm2 ) vs. contralateral (arm = 9.4 ± 3.9 mm2 ; torso = 9.1 ± 4.6 mm2 ) side; this difference was significant both for the arm (P = 0.014) and torso (P = 0.025). Affected (arm: P = 0.010; torso: P = 0.016) but not contralateral (arm: P = 0.42; torso: P = 0.71) vessel areas were significantly elevated compared with control values. Lymphatic cross-sectional areas reduced following MLD on the affected side (pre-MLD: arm = 8.8 ± 1.8 mm2 ; torso = 31.4 ± 26.0 mm2 ; post-MLD: arm = 6.6 ± 1.8 mm2 ; torso = 23.1 ± 24.3 mm2 ). This change was significant in the torso (P = 0.036). The categorical scoring was found to be less specific for detecting lateralizing disease compared to lymphatic-vessel areas.
CONCLUSION
A 3.0T lymphangiography sequence is proposed, which allows for upper extremity lymph stasis to be detected in ∼10 minutes without exogenous contrast agents.
LEVEL OF EVIDENCE: 1 Technical Efficacy: Stage 3 J. Magn. Reson. Imaging 2017;46:1349-1360.