Periodic verification of results' comparability between several analyzers: experience in the application of the EP31-A-IR guideline.

Guiñón Leonor, Illana Francisco J, Cuevas Biel, Canyelles Marina, Martínez-Bru Cecília, García-Osuna Álvaro

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To assess the usefulness of the EP31-A-IR guideline published by the Clinical and Laboratory Standards Institute (CLSI) to perform the periodic verification of results' comparability between several analyzers.

METHODS

Twenty-four biochemistry parameters that could be measured in different analyzers were included: albumin, alkaline phosphatase, alanine aminotransferase, amylase, aspartate aminotransferase, calcium, chloride, C-reactive protein, creatine kinase, creatinine, direct bilirubin, gamma glutamyl transferase, glucose, lactate dehydrogenase, magnesium, phosphate, potassium, sodium, total bilirubin, total cholesterol, total protein, triglycerides, urea and uric acid. In accordance with the EP31-A-IR guideline: (1) Patient samples were selected considering the concentration or activity of interest. (2) Acceptance criteria were established specifically for each concentration or activity level. A quality specification based on biological variation or on state of the art was selected, considering the analytical performance of the available technology. (3) Maximum allowable differences (MAD) between analyzers were calculated. (4) Measurements were performed as stated in appendix B of the guideline. (5) Maximum differences between analyzers were calculated. Results were considered comparable when the maximum difference was less than or equal to the MAD.

RESULTS

For the 24 parameters evaluated, any difference between analyzers exceeded the MAD.

CONCLUSIONS

The EP31-A-IR guideline proved to be useful for periodic verification of results' comparability. However, it must be considered that, to be practicable, it may require to adjust the acceptance criteria in accordance to the analytical performance of the available technology; as well as the number of analytical measurements conforming to the laboratory resources.

Для цитирования: Guiñón Leonor, Illana Francisco J, Cuevas Biel, Canyelles Marina, Martínez-Bru Cecília, García-Osuna Álvaro. Periodic verification of results' comparability between several analyzers: experience in the application of the EP31-A-IR guideline.. Clinical chemistry and laboratory medicine. 2023. DOI: 10.1515/cclm-2023-0994.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.