Hemolysis: a harbinger of adverse outcome after left ventricular assist device implant.

Cowger Jennifer A, Romano Matthew A, Shah Palak, Shah Neha, Mehta Vivek, Haft Jonathan W, Aaronson Keith D, Pagani Francis D

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The clinical relevance of elevated serum markers of hemolysis during left ventricular assist device (LVAD) support has not been fully ascertained.

METHODS

Lactate dehydrogenase (LDH) and serum free hemoglobin (sfHg) values were tallied monthly in 182 patients on HeartMate II (Thoratec, Pleasanton, CA) LVAD support. Peak values for each marker were identified, and 2 hemolysis definitions were applied to the cohort: Hemolysis according to Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) criteria (sfHg > 40 mg/dl with signs/symptoms) and/or hemolysis defined by an LDH ≥ 600 IU/liter (2.5-times the upper limit of laboratory normal). Kaplan-Meier survival free from death, urgent United Network of Organ Sharing 1A transplant for thrombosis, device exchange for thrombosis, and stroke/peripheral embolism was estimated, and Cox hazard ratios (HR) with the 95% confidence interval (95% CI) were calculated. Areas under the receiver-operating characteristic curves (AUCs) for predicting 1-year event-free survival were calculated.

RESULTS

Hemolysis occurred in 32 patients (18%) by INTERMACS criteria and in 68 (37%) patients by LDH criteria. Over a median (25(th), 75(th)) support of 427 days (245, 793 days), there were 78 events. One year event-free survival after the onset of INTERMACS-defined hemolysis was 16% ± 8.3% compared with 85% ± 3.2% in non-hemolyzers (HR, 14.7; 95% CI, 7.9-27; AUC 0.70 ± 0.05; p < 0.001; ). One year event-free survival after the onset of LDH-defined hemolysis was 32% ± 7.2% compared with 89% ± 3.2% in those with persistent LDH values < 600 IU/liter (HR, 8.0; 95% CI, 4.4-14; AUC 0.87 ± 0.04; p < 0.001). Patients who met the LDH hemolysis definition had longer times from hemolysis onset to clinical events and larger magnitudes of risk for embolism and device exchange for thrombosis than those with INTERMACS hemolysis.

CONCLUSIONS

Serum hemolysis marker elevations are associated with increased events in LVAD patients. LDH monitoring provides an earlier diagnosis of adverse events than sfHg, supporting need for a new INTERMACS definition of VAD-associated hemolysis.

Для цитирования: Cowger Jennifer A, Romano Matthew A, Shah Palak, Shah Neha, Mehta Vivek, Haft Jonathan W, Aaronson Keith D, Pagani Francis D. Hemolysis: a harbinger of adverse outcome after left ventricular assist device implant.. The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation. 2013. DOI: 10.1016/j.healun.2013.08.021.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.