MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/jcn.0000000000000114

Медицинская библиотека · The Journal of cardiovascular nursing · 2014

Novel approach to cardiac alarm management on telemetry units.

Whalen Deborah A, Covelle Patricia M, Piepenbrink James C, Villanova Karen L, Cuneo Charlotte L, Awtry Eric H

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

General medical-surgical units struggle with how best to use cardiac monitor alarms to alert nursing staff to important abnormal heart rates (HRs) and rhythms while limiting inappropriate and unnecessary alarms that may undermine both patient safety and quality of care. When alarms are more often false than true, the nursing staff's sense of urgency in responding to alarms is diminished. In this syndrome of "clinical alarm fatigue," the simple burden of alarms desensitizes caregivers to alarms. Noise levels associated with frequent alarms may also heighten patient anxiety and disrupt their perception of a healing environment. Alarm fatigue experienced by nurses and patients is a significant problem and innovative solutions are needed.

OBJECTIVE

The purpose of this quality improvement study was to determine variables that would safely reduce noncritical telemetry and monitor alarms on a general medical-surgical unit where standard manufacturer defaults contributed to excessive audible alarms.

METHODS

Mining of alarm data and direct observations of staff's response to alarms were used to identify the self-reset warning alarms for bradycardia, tachycardia, and HR limits as the largest contributors of audible alarms. In this quality improvement study, the alarms for bradycardia, tachycardia, and HR limits were changed to "crisis," requiring nursing staff to view and act on the alarm each time it sounded. The limits for HR were HR low 45 bpm and HR high 130 bpm.

RESULTS

An overall 89% reduction in total mean weekly audible alarms was achieved on the pilot unit (t = 8.84; P < .0001) without requirement for additional resources or technology. Staff and patient satisfaction also improved. There were no adverse events related to missed cardiac monitoring events, and the incidence of code blues decreased by 50%.

CONCLUSIONS

Alarms with self-reset capabilities may result in an excess number of audible alarms and clinical alarm fatigue. By eliminating self-resetting alarms, the volume of audible alarms and associated clinical alarm fatigue can be significantly reduced without requiring additional resources or technology or compromising patient safety and lead to improvement in both staff and patient satisfaction.

Для цитирования: Whalen Deborah A, Covelle Patricia M, Piepenbrink James C, Villanova Karen L, Cuneo Charlotte L, Awtry Eric H. Novel approach to cardiac alarm management on telemetry units.. The Journal of cardiovascular nursing. 2014. DOI: 10.1097/jcn.0000000000000114.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.