MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3389/fcvm.2022.832242

Медицинская библиотека · Frontiers in cardiovascular medicine · 2022

Real-World Experience With a Large Bore Vascular Closure Device During TAVI Procedure: Features and Predictors of Access-Site Vascular Complications.

Masiero Giulia, D'Angelo Livio, Fovino Luca Nai, Fabris Tommaso, Cardaioli Francesco, Rodinò Giulio, Benedetti Alice, Boiago Mauro, Continisio Saverio, Montonati Carolina, Sciarretta Tommaso, Zuccarelli Vittorio, Scotti Andrea, Lorenzoni Giulia, Pavei Andrea, Napodano Massimo, Fraccaro Chiara, Iliceto Sabino, Marchese Alfredo, Esposito Giovanni, Tarantini Giuseppe

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUNDS: Among vascular closure devices (VCDs), the novel collagen plug-based MANTA VCD is the first designed for large bore percutaneous access. We aimed to assess the features and predictors of access-site vascular complications in an unselected trans-femoral transcatheter aortic valve replacement (TF-TAVR) population.

METHODS

Patients undergoing large bore arteriotomy closure with 18F MANTA VCD following TF-TAVR at a large tertiary care center from September 2019 to January 2021 were prospectively analyzed. Primary Outcome was the MANTA VCD access-site-related complications according to Valve Academic Research Consortium-3 (VARC) definitions. Its incidence and predictors were evaluated.

RESULTS

Eighty-eight patients (median age 82 years, 48% male, 3.3 median Society of Thoracic Surgeons score) undergoing TF-TAVR were included, mostly (63%) treated with a self-expandable device and with outer diameter sizes varied from 18F to 24-F. MANTA VCD technical success rate was 98%, while 10 patients (11%) experienced MANTA VCD access-site vascular complications which included 8% of minor complications and only to 2% of major events resulting in VARC type ≥2 bleeding. Vessel occlusion/stenosis (60%), perforation (20%), and pseudoaneurysm/dissection/hematoma (20%) occurred, but all were managed without surgical treatment. Independent predictors of failure were age (p = 0.04), minimum common femoral artery diameter (CFA) (p < 0.01), sheath-to-femoral-artery ratio (SFAR) (p < 0.01), and a lower puncture height (p = 0.03). A CFA diameter <7.1 mm with a SFAR threshold of 1.01 were associated with VCD failure.

CONCLUSIONS

In a more comers TF-TAVR population, MANTA VCD was associated with reassuring rates of technical success and major access-site vascular complications. Avoiding lower vessel size and less puncture site distance to CFA bifurcation might further improve outcomes.

Для цитирования: Masiero Giulia, D'Angelo Livio, Fovino Luca Nai, Fabris Tommaso, Cardaioli Francesco, Rodinò Giulio, Benedetti Alice, Boiago Mauro, Continisio Saverio, Montonati Carolina, Sciarretta Tommaso, Zuccarelli Vittorio, Scotti Andrea, Lorenzoni Giulia, Pavei Andrea, Napodano Massimo, Fraccaro Chiara, Iliceto Sabino, Marchese Alfredo, Esposito Giovanni, Tarantini Giuseppe. Real-World Experience With a Large Bore Vascular Closure Device During TAVI Procedure: Features and Predictors of Access-Site Vascular Complications.. Frontiers in cardiovascular medicine. 2022. DOI: 10.3389/fcvm.2022.832242.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.