Accuracy of indirect bonding trays - a measurement algorithm.
Аннотация
AIM: To present an image-processing measurement algorithm to evaluate the transfer accuracy of indirect bonding (IDB) trays, exemplified by a CAD/CAM-based IDB tray integrated into a digital orthodontic workflow.
MATERIALS AND
METHODS
Plaster casts of 24 patients with full dentition and different malocclusions were scanned with an intraoral scanner (Trios; 3Shape, Copenhagen, Denmark) to obtain digital models, which served for the virtual placement of orthodontic brackets in simulation software (OnyxCeph; Image Instruments, Chemnitz, Germany). The resulting STL files were sent to a dental laboratory (CA Digital; Hilden, Germany) for the production of INDIVIDUA IDB trays. These trays were used to transfer the brackets to the respective plaster casts. Finally, a second scan was performed to record the actual bracket positions. The transfer accuracy was then analyzed by a measurement algorithm scripted to automation, which calculated the deviations of the planned and real bracket positions with a local best-fit alignment, resulting in three linear and three angular measurements for each bracket.
RESULTS
In total, 622 brackets and tubes were transferred successfully. The presented algorithm analyzed the transfer accuracy and demonstrated that the linear measurements were 98.3% within the range of the American Board of Orthodontics standard. The angular measurements were 86.7% within this range when the INDIVIDUA IDB tray was used.
CONCLUSION
Scripted measurement algorithms facilitated the evaluation of present and future materials and designs for IDB trays to obtain an efficient solution for orthodontic practice. The INDIVIDUA IDB tray is a digital alternative to conventional IDB trays (Int J Comput Dent 2022;25(3):295-302; doi: 10.3290/j.ijcd.b2599775).
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.