JACC. Clinical electrophysiology · 2021
Radiofrequency Energy Applications Targeting Significant Residual Leaks After Watchman Implantation: A Prospective, Multicenter Experience.
СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Della Rocca Domenico G, Murtaza Ghulam, Di Biase Luigi, Akella Krishna, Krishnan Subramaniam C, Magnocavallo Michele, Mohanty Sanghamitra, Gianni Carola, Trivedi Chintan, Lavalle Carlo, Forleo Giovanni B, Natale Veronica N, Tarantino Nicola, Romero Jorge, Gopinathannair Rakesh, Patel Philip J, Bassiouny Mohamed, Del Prete Armando, Al-Ahmad Amin, Burkhardt J David, Gallinghouse G Joseph, Sanchez Javier E, Doshi Shephal K, Horton Rodney P, Lakkireddy Dhanunjaya, Natale Andrea
Аннотация
OBJECTIVES: The aim of this study was to evaluate the efficacy of radiofrequency (RF) energy applications targeting the atrial side of a significant residual leak in patients with acute and chronic evidence of incomplete percutaneous left atrial appendage (LAA) occlusion.
BACKGROUND: RF applications have been proved to prevent recanalization of intracranial aneurysms after coil embolization, thereby favoring complete sealing. From a mechanistic standpoint, in vitro and in vivo experiments have demonstrated that RF promotes collagen deposition and tissue retraction.
METHODS: Forty-three patients (mean age 75 ± 7 years mean CHA2DS2-VASc score 4.6 ± 1.4, mean HAS-BLED score 4.0 ± 1.1) with residual leaks ≥4 mm after Watchman implantation were enrolled. Procedural success was defined as complete LAA occlusion or presence of a mild or minimal (1- to 2-mm) peridevice leak on follow-up transesophageal echocardiography (TEE), which was performed approximately 45 days after the procedure.
RESULTS: RF-based leak closure was performed acutely after Watchman implantation in 19 patients (44.2%) or scheduled after evidence of significant leaks on follow-up TEE in 24 others (55.8%). The median leak size was 5 mm (range: 4-7 mm). On average, 18 ± 7 RF applications per patient (mean maximum contact force 16 ± 3 g, mean power 44 ± 2 W, mean RF time 5.1 ± 2.5 minutes) were performed targeting the atrial edge of the leak. Post-RF median leak size was 0 mm (range: 0-1 mm). A very low rate (2.3% [n = 1]) of major periprocedural complications was observed. Follow-up TEE revealed complete LAA sealing in 23 patients (53.5%) and negligible residual leaks in 15 (34.9%).
CONCLUSIONS: RF applications targeting the atrial edge of a significant peri-Watchman leak may promote LAA sealing via tissue remodeling, without increasing complications. (RF Applications for Residual LAA Leaks [REACT]; NCT04726943).
BACKGROUND: RF applications have been proved to prevent recanalization of intracranial aneurysms after coil embolization, thereby favoring complete sealing. From a mechanistic standpoint, in vitro and in vivo experiments have demonstrated that RF promotes collagen deposition and tissue retraction.
METHODS: Forty-three patients (mean age 75 ± 7 years mean CHA2DS2-VASc score 4.6 ± 1.4, mean HAS-BLED score 4.0 ± 1.1) with residual leaks ≥4 mm after Watchman implantation were enrolled. Procedural success was defined as complete LAA occlusion or presence of a mild or minimal (1- to 2-mm) peridevice leak on follow-up transesophageal echocardiography (TEE), which was performed approximately 45 days after the procedure.
RESULTS: RF-based leak closure was performed acutely after Watchman implantation in 19 patients (44.2%) or scheduled after evidence of significant leaks on follow-up TEE in 24 others (55.8%). The median leak size was 5 mm (range: 4-7 mm). On average, 18 ± 7 RF applications per patient (mean maximum contact force 16 ± 3 g, mean power 44 ± 2 W, mean RF time 5.1 ± 2.5 minutes) were performed targeting the atrial edge of the leak. Post-RF median leak size was 0 mm (range: 0-1 mm). A very low rate (2.3% [n = 1]) of major periprocedural complications was observed. Follow-up TEE revealed complete LAA sealing in 23 patients (53.5%) and negligible residual leaks in 15 (34.9%).
CONCLUSIONS: RF applications targeting the atrial edge of a significant peri-Watchman leak may promote LAA sealing via tissue remodeling, without increasing complications. (RF Applications for Residual LAA Leaks [REACT]; NCT04726943).