OBJECTIVES

This study sought to investigate patient intermediate-term outcomes after transcatheter pulmonary valve replacement (TPVR) with Edwards SAPIEN valve.

BACKGROUND

The Edwards SAPIEN valve, initially designed for percutaneous aortic valve replacement, has been approved for TPVR in patients with dysfunctional right ventricular outflow tracts (RVOT), but only short-term follow-up has been reported.

METHODS

From 2011 to 2016, 62 patients undergoing successful TPVR using the SAPIEN XT valve were consecutively included into the study. Primary efficacy and safety endpoints were defined as freedom from valve-reintervention and freedom from infective endocarditis at last follow-up, respectively.

RESULTS

The primary efficacy outcome was met for 87.1% patients after a mean follow-up of 4.6 ± 1.8 years, corresponding to a freedom of reintervention at 5 years of 89% (95% CI 74.8-95.6%). Reinterventions were exclusively due to recurrent obstruction, no significant valvular regurgitation was observed. One case of infective endocarditis was reported, corresponding to a rate of 0.35% per patient-year (95% CI 0.01-2.00%). At 5 years, freedom from infective endocarditis was 98.4% (95% CI 89.1-99.8%). Six patients died or were transplanted due to advanced cardiac failure, without relationship with TPVR. In univariate analysis, reintervention was associated with young age, a smaller tube-graft, a higher pulmonary valve gradient after the procedure and a ratio of largest implanted stent diameter to invasive balloon conduit diameter over 1.35.

CONCLUSIONS

This study documents the mid-term safety and efficacy of the Edwards SAPIEN XT valve in patients with dysfunctional RVOT, and identifies a patient profile associated with an uncertain benefit-risk balance.

Для цитирования: Le Ruz Robin, Plessis Julien, Houeijeh Ali, Baruteau Alban-Elouen, Le Gloan Laurianne, Warin Fresse Karine, Karsenty Clément, Petit Jérôme, Godart François, Hascoët Sébastien, Guérin Patrice. Edwards SAPIEN XT transcatheter pulmonary valve implantation: 5-year follow-up in a French Registry.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2021. DOI: 10.1002/ccd.29862.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.