DOI 10.1007/s12328-021-01448-w
Swallow-induced syncope after magnetic sphincter augmentation: a case report and physiologic explanation.
Аннотация
Swallow-induced syncope is a rare cause of syncope that occurs during or immediately after swallowing. This phenomenon has been reported in association with few esophageal pathologies and the likely explanation is a vagal reflex during deglutition that results in inhibition of the cardiac conduction system. This report describes a case of swallow-induced syncope related to the implantation of a magnetic sphincter augmenting (MSA) device. Two episodes of syncope after food bolus occurred with the device in place and upon removal of the device, the patient had no further episodes of syncope. Vagal stimulation from distention of the esophagus or vagus nerve contact irritation by the implant are the potential explanation for syncope in this patient. Although this is an extremely rare complication of magnetic sphincter device augmentation, it is one that physicians should be cognizant of given the dangers of syncope events.