DOI 10.1016/j.athoracsur.2013.06.089

Медицинская библиотека · The Annals of thoracic surgery · 2013

Intermediate follow-up results from the multicenter engager European pivotal trial.

Holzhey David, Linke Axel, Treede Hendrik, Baldus Stephan, Bleiziffer Sabine, Wagner Anke, Börgermann Jochen, Scholtz Werner, Vanoverschelde Jean-Louis, Falk Volkmar

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Optimal transcatheter aortic valve (TAVI) results require accurate valve positioning, including anatomically correct orientation and secure fixation within the aortic annulus, thereby potentially decreasing paravalvular regurgitation. The new Engager (Medtronic 3F Therapeutics, Santa Ana, CA) transapical valve system captures the native leaflets for sealing and allows for tactile feedback during valve placement. We report initial safety and performance outcomes of the Engager system through 6 months in patients with severe aortic valve stenosis at high risk for surgical aortic valve replacement.

METHODS

An interim analysis was performed on the first 61 enrolled September 2011 through May 2012. Inclusion criteria comprised severe aortic stenosis, New York Heart Association functional class of II or greater, logistic European System for Cardiac Operative Risk Evaluation (EuroSCORE) of 20% or greater, or contraindication to surgical aortic valve replacement. The primary endpoint was all-cause mortality at 30 days. Patients were evaluated 24 to 48 hours post-procedure, at hospital discharge, 30 days and 6 months. Follow-up is planned annually through five years.

RESULTS

Baseline characteristics for the 61 patients were mean age 81.9 ± 4.4 years, 62.3% female, 88.5% New York Heart Association class III/IV, 52.5% coronary artery disease, and 54.2% extracardiac arteriopathy. For all of the attempted implantations (n = 60), the Engager prosthesis was positioned in the correct anatomic position without conversions to surgery, second valve implantation, device malposition, aortic annular rupture, or coronary obstruction. All-cause mortality was 9.9% at 30 days and 16.9% at 6 months. The baseline mean aortic valve gradient was 43.7 ± 16.7 mm Hg and 11.5 ± 5.0 mm Hg at 30 days, and showed similar reduction at 6 months (13.9 ± 6.2 mm Hg). There was no paravalvular regurgitation greater than mild through 6 months.

CONCLUSIONS

Early postoperative results support implantation success and valve safety. Analysis for 6 month outcomes shows stable hemodynamic performance and clinical outcome. (Transapical Implantation of the Medtronic Engager Transcatheter Aortic Valve Implantation System-the Engager European Pivotal Trial; NCT01348438).

Для цитирования: Holzhey David, Linke Axel, Treede Hendrik, Baldus Stephan, Bleiziffer Sabine, Wagner Anke, Börgermann Jochen, Scholtz Werner, Vanoverschelde Jean-Louis, Falk Volkmar. Intermediate follow-up results from the multicenter engager European pivotal trial.. The Annals of thoracic surgery. 2013. DOI: 10.1016/j.athoracsur.2013.06.089.

Предметные рубрики MeSH