DOI 10.1016/j.phro.2023.100505

Медицинская библиотека · Physics and imaging in radiation oncology · 2023

Accurate, repeatable, and geometrically precise diffusion-weighted imaging on a 0.35 T magnetic resonance imaging-guided linear accelerator.

Weygand Joseph, Armstrong Tess, Bryant John Michael, Andreozzi Jacqueline M, Oraiqat Ibrahim M, Nichols Steven, Liveringhouse Casey L, Latifi Kujtim, Yamoah Kosj, Costello James R, Frakes Jessica M, Moros Eduardo G, El Naqa Issam M, Naghavi Arash O, Rosenberg Stephen A, Redler Gage

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: Diffusion weighted imaging (DWI) allows for the interrogation of tissue cellularity, which is a surrogate for cellular proliferation. Previous attempts to incorporate DWI into the workflow of a 0.35 T MR-linac (MRL) have lacked quantitative accuracy. In this study, accuracy, repeatability, and geometric precision of apparent diffusion coefficient (ADC) maps produced using an echo planar imaging (EPI)-based DWI protocol on the MRL system is illustrated, and in vivo potential for longitudinal patient imaging is demonstrated.

MATERIALS AND

METHODS

Accuracy and repeatability were assessed by measuring ADC values in a diffusion phantom at three timepoints and comparing to reference ADC values. System-dependent geometric distortion was quantified by measuring the distance between 93 pairs of phantom features on ADC maps acquired on a 0.35 T MRL and a 3.0 T diagnostic scanner and comparing to spatially precise CT images. Additionally, for five sarcoma patients receiving radiotherapy on the MRL, same-day in vivo ADC maps were acquired on both systems, one of which at multiple timepoints.

RESULTS

Phantom ADC quantification was accurate on the 0.35 T MRL with significant discrepancies only seen at high ADC. Average geometric distortions were 0.35 (±0.02) mm and 0.85 (±0.02) mm in the central slice and 0.66 (±0.04) mm and 2.14 (±0.07) mm at 5.4 cm off-center for the MRL and diagnostic system, respectively. In the sarcoma patients, a mean pretreatment ADC of 910x10-6 (±100x10-6) mm2/s was measured on the MRL.

CONCLUSIONS

The acquisition of accurate, repeatable, and geometrically precise ADC maps is possible at 0.35 T with an EPI approach.

Для цитирования: Weygand Joseph, Armstrong Tess, Bryant John Michael, Andreozzi Jacqueline M, Oraiqat Ibrahim M, Nichols Steven, Liveringhouse Casey L, Latifi Kujtim, Yamoah Kosj, Costello James R, Frakes Jessica M, Moros Eduardo G, El Naqa Issam M, Naghavi Arash O, Rosenberg Stephen A, Redler Gage. Accurate, repeatable, and geometrically precise diffusion-weighted imaging on a 0.35 T magnetic resonance imaging-guided linear accelerator.. Physics and imaging in radiation oncology. 2023. DOI: 10.1016/j.phro.2023.100505.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.