Swan Marc C, Bucknall David G, Czernuszka Jan T, Pigott David W, Goodacre Timothy E E
Аннотация
BACKGROUND: The advent of self-inflating hydrogel tissue expanders heralded a significant advance in the reconstructive potential of this technique. Their use, however, is limited by their uncontrolled isotropic (i.e., uniform in all directions) expansion.
METHODS: Anisotropy (i.e., directional dependence) was achieved by annealing a hydrogel copolymer of poly(methyl methacrylate-co-vinyl pyrrolidone) under a compressive load for a specified time period. The expansion ratio is dictated by the percentage of vinyl pyrrolidone content and the degree of compression. The expansion rate is modified by incorporating the polymer within a silicone membrane. The in vivo efficacy of differing prototype devices was investigated in juvenile pigs under United Kingdom Home Office Licence. The devices were implanted within a submucoperiosteal pocket in a total of six porcine palates; all were euthanized by 6 weeks after implantation. A longitudinal volumetric assessment of the expanded tissue was conducted, in addition to postmortem analysis of the bony and mucoperiosteal palatal elements.
RESULTS: Uncoated devices caused excessive soft-tissue expansion that resulted in mucoperiosteal ulceration, thus necessitating animal euthanasia. The silicone-coated devices produced controlled soft-tissue expansion over the 6-week study period. There was a statistically significant increase in the volume of expanded soft tissue with no evidence of a significant acute inflammatory response to the implant, although peri-implant capsule formation was observed. Attenuation of the bony palatal shelf was noted.
CONCLUSION: A unique anisotropic hydrogel device capable of controlled expansion has been developed that addresses a number of the shortcomings of the technology hitherto available.
METHODS: Anisotropy (i.e., directional dependence) was achieved by annealing a hydrogel copolymer of poly(methyl methacrylate-co-vinyl pyrrolidone) under a compressive load for a specified time period. The expansion ratio is dictated by the percentage of vinyl pyrrolidone content and the degree of compression. The expansion rate is modified by incorporating the polymer within a silicone membrane. The in vivo efficacy of differing prototype devices was investigated in juvenile pigs under United Kingdom Home Office Licence. The devices were implanted within a submucoperiosteal pocket in a total of six porcine palates; all were euthanized by 6 weeks after implantation. A longitudinal volumetric assessment of the expanded tissue was conducted, in addition to postmortem analysis of the bony and mucoperiosteal palatal elements.
RESULTS: Uncoated devices caused excessive soft-tissue expansion that resulted in mucoperiosteal ulceration, thus necessitating animal euthanasia. The silicone-coated devices produced controlled soft-tissue expansion over the 6-week study period. There was a statistically significant increase in the volume of expanded soft tissue with no evidence of a significant acute inflammatory response to the implant, although peri-implant capsule formation was observed. Attenuation of the bony palatal shelf was noted.
CONCLUSION: A unique anisotropic hydrogel device capable of controlled expansion has been developed that addresses a number of the shortcomings of the technology hitherto available.