DOI 10.1016/s0735-1097(98)00245-9
Remodeling after directional coronary atherectomy (with and without adjunct percutaneous transluminal coronary angioplasty): a serial angiographic and intravascular ultrasound analysis from the Optimal Atherectomy Restenosis Study.
Аннотация
OBJECTIVES
The intravascular ultrasound (IVUS) substudy of OARS (Optimal Atherectomy Restenosis Study) was designed to assess the mechanisms of restenosis after directional coronary atherectomy (DCA).
BACKGROUND
Recent serial IVUS studies have indicated that late lumen loss after interventional procedures was determined primarily by the direction and magnitude of arterial remodeling, not by cellular proliferation.
METHODS
Complete quantitative coronary angiography (QCA) and IVUS were obtained in 104 patients before and after intervention and during follow-up. All studies were performed after administration of 200 microg of intracoronary nitroglycerin. Angiographic measurements included minimum lumen diameter (MLD), interpolated reference diameter and diameter stenosis (DS). Intravascular ultrasound measurements included lesion and reference external elastic membrane (EEM), lumen and plaque+media cross-sectional area (CSA). The axial location of the lesion site was at the smallest follow-up lumen CSA; the reference segment was the most normal-looking cross section within 10 mm proximal to the lesion but distal to any major side branch. Results are reported as mean +/- one standard deviation.
RESULTS
The QCA reference decreased from 3.51 +/- 0.46 mm to 3.22 +/- 0.44 mm; the MLD decreased from 3.22 +/- 0.47 mm to 2.03 +/- 0.72 mm; and the DS increased from 8 +/- 10% to 38 +/- 20%. On IVUS, the decrease in lumen CSA (from 8.8 +/- 2.5 mm2 to 5.5 +/- 4.0 mm2) was associated with a significant decrease in EEM (from 19.7 +/- 5.6 mm2 to 16.9 +/- 6.2 mm2); there was no significant increase in P+M (from 10.9 +/- 4.2 mm2 to 11.3 +/- 3.9 mm2). A change in lumen correlated with a change in EEM (r = 0.790, p < 0.0001), not with a change in P+M (r = 0.133, p = 0.2258). A decrease in reference EEM (from 19.1 +/- 7.7 mm2 to 17.6 +/- 8.0 mm2) also correlated with a decrease in lesion EEM (r = 0.665, p < 0.0001). Results in restenotic lesions were similar.
CONCLUSION
Restenosis after optimal DCA is caused primarily by a decrease in EEM CSA that extends into contiguous reference segments.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.