Sources of error in intra-arterial pressure measurements across a stenosis.
Аннотация
OBJECTIVE
To investigate potential errors associated with different techniques of intra-arterial pressure measurement at angiography.
MATERIALS AND
METHODS
An experimental model of an arterial stenosis was developed. Experiments were performed to assess the relevance of catheter position, catheter direction and catheter type on the recorded intraluminal pressure. Transstenotic pressure gradients were recorded with and without angiographic catheters crossing the stenosis.
RESULTS
At physiological flow rates angiographic catheter type does not influence the recorded pressure. At high flow rates through tight stenoses there is a significant catheter-related difference in recorded pressure adjacent to a stenosis. Downstream pressures may be altered by up to 85 mmHg when standard angiographic catheters are placed across a stenosis.
CONCLUSION
The different techniques employed to measure pressure differences across a model stenosis may introduce significant errors up to 85 mmHg. Care must be taken when pressure measurements alone are used to interpret the clinical significance of a stenosis. In low flow conditions there may not be a detectable pressure gradient across a 95% stenosis.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.