Аннотация

PURPOSE: The objective of this paper was to provide an update on the clinical and experimental evaluation of radioactive stents for the prevention of restenosis.

MATERIALS AND

METHODS

Direct ion implantation of 32P onto the surface of a 15-mm length balloon expandable stainless-steel Palmaz-Schatz stent was employed to render this commercially available vascular stent radioactive. 32Phosphorous, a pure beta-particle-emitting radioisotope, was selected because of its short half-life (14.3 days) and limited range of tissue penetration (3-4 mm). The vascular response to radioactive 7-mm length Palmaz-Schatz stents with activities 0.14 to 23 microCi of 32P were evaluated in animal models of arterial injury and restenosis. The Phase-1 isostent for restenosis intervention study (IRIS trial) was a nonrandomized safety trial designed to evaluate the use of a low activity 32P (0.5 to 1.5 microCi) 15-mm length Palmaz-Schatz stent for the treatment of de novo or restenosis native coronary arterial lesions.

RESULTS

In the porcine coronary restenosis model, at < or =0.5 microCi and > or =3.0 microCi stent activities, there was a 30% reduction in the neointimal and percent area stenosis as compared to nonradioactive stents. The 1.0 microCi stents, however, had nearly 2-fold greater neointimal formation and more luminal narrowing than the control stents. In the Phase 1 IRIS trial, 57 patients with symptomatic de novo or restenosis native coronary lesions have been treated with low activity (0.5 to 1.5 microCi) 32P Palmaz-Schatz coronary stents. Fifty-seven stents were successfully implanted without a major procedural complication (death, urgent coronary bypass, Q-wave myocardial infarction). There were no cases of stent thrombosis, target vessel revascularization, or other adverse events in the first 30 days after implant.

CONCLUSION

The early clinical results with a low-activity 32P Palmaz-Schatz radioactive stent demonstrate sufficient procedural and 30-day event-free survival to warrant consideration of additional clinical studies to determine the safety and efficacy of this therapy for the prevention of restenosis. Future studies will focus on optimal stent design for delivery of radiation, and will further evaluate safe and effective dosing strategies.

Для цитирования: Carter A J, Fischell T A. Current status of radioactive stents for the prevention of in-stent restenosis.. International journal of radiation oncology, biology, physics. 1998. DOI: 10.1016/s0360-3016(98)00024-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.