DOI 10.1016/s0003-9993(00)90215-6

Медицинская библиотека · Archives of physical medicine and rehabilitation · 2000

Bladder volume determination: portable 3-D versus stationary 2-D ultrasound device.

Schnider P, Birner P, Gendo A, Ratheiser K, Auff E

Клиническое исследованиеСравнительное исследованиеКонтролируемое клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To investigate how accurately a portable three-dimensional (3-D) scanner and a multipurpose two-dimensional (2-D) real-time scanner determined bladder volumes.

STUDY

DESIGN

Prospective, controlled clinical trial, single-blind, crossover design.

SETTING AND

PARTICIPANTS

Twenty-three inpatients with permanent bladder catheters participated voluntarily in this study.

METHODS

The bladders of 20 patients were filled through an indwelling catheter with 60, 110, 160, 210, and 260 mL sterile normal saline. Volumes were measured twice with each device. Measurements were compared with the actual bladder volumes.

RESULTS

The 2-D device showed better reproducibility, particularly at lower bladder volumes. The 3-D scanner showed a significant difference between the two measurements at 160 mL (p<.05) and had poor reproducibility at 110, 210, and 260 mL. Both devices overestimated actual bladder volume at fillings of <160 mL and underestimated it at fillings of > or =160 mL. The range between the 25th and 75th percentiles was always larger for the 3-D scanner, except for the 210 mL reading.

CONCLUSION

Both devices showed sufficient accuracy for clinical practice. Ultrasound measurements of >110 mL should be followed by catheterization to detect potentially harmful bladder volumes.

Для цитирования: Schnider P, Birner P, Gendo A, Ratheiser K, Auff E. Bladder volume determination: portable 3-D versus stationary 2-D ultrasound device.. Archives of physical medicine and rehabilitation. 2000. DOI: 10.1016/s0003-9993(00)90215-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.