MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0139973

Медицинская библиотека · PloS one · 2015

Accuracy of Continuous Glucose Monitoring Measurements in Normo-Glycemic Individuals.

Akintola Abimbola A, Noordam Raymond, Jansen Steffy W, de Craen Anton J, Ballieux Bart E, Cobbaert Christa M, Mooijaart Simon P, Pijl Hanno, Westendorp Rudi G, van Heemst Diana

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The validity of continuous glucose monitoring (CGM) is well established in diabetic patients. CGM is also increasingly used for research purposes in normo-glycemic individuals, but the CGM validity in such individuals is unknown. We studied the accuracy of CGM measurements in normo-glycemic individuals by comparing CGM-derived versus venous blood-derived glucose levels and measures of glycemia and glycemic variability.

METHODS

In 34 healthy participants (mean age 65.7 years), glucose was simultaneously measured every 10 minutes, via both an Enlite® CGM sensor, and in venous blood sampled over a 24-hour period. Validity of CGM-derived individual glucose measurements, calculated measures of glycemia over daytime (09:00h-23:00h) and nighttime (23:00h-09:00h), and calculated measures of glycemic variability (e.g. 24h standard deviation [SD]) were assessed by Pearson correlation coefficients, mean absolute relative difference (MARD) and paired t-tests.

RESULTS

The median correlation coefficient between CGM and venous glucose measurements per participant was 0.68 (interquartile range: 0.40-0.78), and the MARD was 17.6% (SD = 17%). Compared with venous sampling, the calculated measure of glycemia during daytime was 0.22 mmol/L higher when derived from CGM, but no difference was observed during nighttime. Most measures of glycemic variability were lower with CGM than with venous blood sampling (e.g., 24h SD: 1.07 with CGM and 1.26 with venous blood; p-value = 0.004).

CONCLUSION

In normo-glycemic individuals, CGM-derived glucose measurements had good agreement with venous glucose levels. However, the measure of glycemia was higher during the day and most measures of glycemic variability were lower when derived from CGM.

Для цитирования: Akintola Abimbola A, Noordam Raymond, Jansen Steffy W, de Craen Anton J, Ballieux Bart E, Cobbaert Christa M, Mooijaart Simon P, Pijl Hanno, Westendorp Rudi G, van Heemst Diana. Accuracy of Continuous Glucose Monitoring Measurements in Normo-Glycemic Individuals.. PloS one. 2015. DOI: 10.1371/journal.pone.0139973.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.