Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2017
Capillary isoelectric focusing with whole column imaging detection (iCIEF): A new approach to the characterization and quantification of salivary α-amylase.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Zarabadi Atefeh S, Huang Tiemin, Mielke John G
Аннотация
Saliva is an easily collected biological fluid with potentially important diagnostic value. While gel electrophoresis is generally used for salivary analysis, we employed the capillary isoelectric focusing technique to allow for a rapid, automated mode of electrophoresis. Capillary isoelectric focusing coupled with UV whole column imaging detection (iCIEF) was used to develop a robust protocol to characterize salivary α-amylase collected from various glands. Notably, three sample preparation methods were examined: ultrafiltration, gel-filtration, and starch affinity interaction with salivary amylase. Salivary α-amylase separated into two major peaks before sample treatment; while both filtration methods and starch affinity interaction of salivary amylase enhanced the resolution of isozymes, desalting with gel-filtration displayed the best recovery and the highest resolution of isozymes. Good agreement existed between the observed isoelectric points and the values reported in the literature. In addition, a high level of precision was apparent, and the relative standard deviation for replicates was less than 0.5% for pIs (peak positions) and below 10% for peak area. Furthermore, saliva secreted from the parotid gland proved to have a higher amylase content compared to either secretions from the submandibular/sublingual complex, or whole saliva, as well as amylase enhancement under stimulation. The results suggest that the iCIEF technique can be used to accurately resolve and quantitate amylase isozymes in a rapid and automated fashion, and that gel-filtration should be applied to saliva samples beforehand to allow for optimal purification and characterization.