MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1001/archsurg.1975.01360170148023

Медицинская библиотека · Archives of surgery (Chicago, Ill. : 1960) · 1975

Mitral valve replacement with the Hancock stabilized glutaraldehyde valve. Clinical and laboratory evaluation.

Buch W S, Pipkin R D, Hancock W D, Fogarty T J

СтатьяПеревод пока не подготовлен

Аннотация

From March 1971 through April 1975, one hundred twenty patients underwent mitral valve replacement with a Hancock "stabilized glutaraldehyde process" porcine aortic xenograft. A simultaneous canine experimental series was also carried out. In the clinical series, the early mortality was 8.3%. Actuarial analyses of all patients predicts survival at two years of 81.0% and at four years of 70.0%. The predicted survival for patients without coronary disease or prior prosthetic valve replacement is 87.5% at two years and 77.5% at four years. There were four thromboembolic episodes, a rate of 2.4% per patient-year. None were fatal. No valve failure were noted. Histologic examination and shrink temperature analysis of recovered valves show excellent tissue preservation at 40 months. The data indicate that the Hancock valve is durable, enjoys a low incidence of thromboembolism, and may be the valve of choice for mitral valve replacement.

Для цитирования: Buch W S, Pipkin R D, Hancock W D, Fogarty T J. Mitral valve replacement with the Hancock stabilized glutaraldehyde valve. Clinical and laboratory evaluation.. Archives of surgery (Chicago, Ill. : 1960). 1975. DOI: 10.1001/archsurg.1975.01360170148023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.