[Evaluation of two arterial closure devices, Angioseal and Perclose, in coronary catheter interventions].
Аннотация
OBJECTIVE
To assess the efficacy and safety of two arterial closure devices, Angioseal and Perclose, in patients undergoing coronary angiography and invasive interventions.
METHODS
From January 2001 to April 2011, 997 inpatients underwent coronary angiography and interventions with arterial closure using Perclose (486 cases) or Angioseal (511 cases). The time to ambulation and hemostasis, major vascular complications and deployment success rate with the two devices were compared.
RESULTS
The time to hemostasis was significantly shorter in Angioseal group than in Perclose group (3∓0.9 min vs 10.8∓4.8 min, P<0.001), but the time to ambulation was comparable between the two groups (6.4∓1.2 h vs 6.3∓0.7 h, P>0.05). The incidences of vascular complications showed no significant differences between the two groups (4.5% vs 3.7%, P>0.05), and none of the cases in either group developed femoral artery thrombosis or low limb embolism following the procedures. The deployment success rate was comparable between the two groups (97.8% vss 98.6%, P>0.05), and deployment failure was associated mainly with mishandling and design defect of the devices.
CONCLUSIONS
Angioseal and Perclose are both effective and safe for arterial closure with reduced hemostasis and ambulation time and low incidences of vascular complications. Angioseal appears to have better performance than Perclose in shortening the hemostasis time and is easier to handle.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.