MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/1532-429x-13-29

Медицинская библиотека · Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2011

Cardiac resynchronization therapy guided by late gadolinium-enhancement cardiovascular magnetic resonance.

Leyva Francisco, Foley Paul W X, Chalil Shajil, Ratib Karim, Smith Russell E A, Prinzen Frits, Auricchio Angelo

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Myocardial scarring at the LV pacing site leads to incomplete resynchronization and a suboptimal symptomatic response to CRT. We sought to determine whether the use of late gadolinium cardiovascular magnetic resonance (LGE-CMR) to guide left ventricular (LV) lead deployment influences the long-term outcome of cardiac resynchronization therapy (CRT).

METHODS

559 patients with heart failure (age 70.4 ± 10.7 yrs [mean ± SD]) due to ischemic or non-ischemic cardiomyopathy underwent CRT. Implantations were either guided (+CMR) or not guided (-CMR) by LGE-CMR prior to implantation. Fluoroscopy and LGE-CMR were used to localize the LV lead tip and and myocardial scarring retrospectively. Clinical events were assessed in three groups: +CMR and pacing scar (+CMR+S); CMR and not pacing scar (+CMR-S), and; LV pacing not guided by CMR (-CMR).

RESULTS

Over a maximum follow-up of 9.1 yrs, +CMR+S had the highest risk of cardiovascular death (HR: 6.34), cardiovascular death or hospitalizations for heart failure (HR: 5.57) and death from any cause or hospitalizations for major adverse cardiovascular events (HR: 4.74) (all P < 0.0001), compared with +CMR-S. An intermediate risk of meeting these endpoints was observed for -CMR, with HRs of 1.51 (P = 0.0726), 1.61 (P = 0.0169) and 1.87 (p = 0.0005), respectively. The +CMR+S group had the highest risk of death from pump failure (HR: 5.40, p < 0.0001) and sudden cardiac death (HR: 4.40, p = 0.0218), in relation to the +CMR-S group.

CONCLUSIONS

Compared with a conventional implantation approach, the use of LGE-CMR to guide LV lead deployment away from scarred myocardium results in a better clinical outcome after CRT. Pacing scarred myocardium was associated with the worst outcome, in terms of both pump failure and sudden cardiac death.

Для цитирования: Leyva Francisco, Foley Paul W X, Chalil Shajil, Ratib Karim, Smith Russell E A, Prinzen Frits, Auricchio Angelo. Cardiac resynchronization therapy guided by late gadolinium-enhancement cardiovascular magnetic resonance.. Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance. 2011. DOI: 10.1186/1532-429x-13-29.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.