Safety and feasibility of a novel adjustable mitral annuloplasty ring: a multicentre European experience.

Andreas Martin, Doll Nicolas, Livesey Steve, Castella Manuel, Kocher Alfred, Casselman Filip, Voth Vladimir, Bannister Christina, Encalada Palacios Juan F, Pereda Daniel, Laufer Guenther, Czesla Markus

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

Recurrent mitral regurgitation is a significant problem after mitral valve repair in patients with functional valve disease. We report the safety and feasibility of a novel adjustable mitral annuloplasty device that permits downsizing of the anterior-posterior diameter late after initial surgery.

METHODS

In this multicentre, non-randomized, observational register, patients with moderate or severe mitral regurgitation undergoing surgical mitral valve repair with the MiCardia EnCorSQ™ Mitral Valve Repair system were evaluated. Patient characteristics, operative specifications and results as well as postoperative follow-up were collected for all five centres.

RESULTS

Ninety-four patients with a median age of 71 (64-75) years (EuroSCORE II 6.7 ± 6.3; 66% male, 48% ischaemic MR, 37% dilated cardiomyopathy and 15% degenerative disease) were included. Operative mortality was 1% and the 1-year survival was 93%. Ring adjustment was attempted in 12 patients at a mean interval of 9 ± 6 months after surgery. In three of these attempts, a technical failure occurred. In 1 patient, mitral regurgitation was reduced two grades, in 2 patients mitral regurgitation was reduced one grade and in 6 patients, mitral regurgitation did not change significantly. The mean grade of mitral regurgitation changed from 2.9 ± 0.9 to 2.1 ± 0.7 (P = 0.02). Five patients were reoperated after 11 ± 9 months (Ring dehiscence: 2; failed adjustment: 3).

CONCLUSION

We conclude that this device may provide an additional treatment option in patients with functional mitral regurgitation, who are at risk for reoperation due to recurrent mitral regurgitation. Clinical results in this complex disease were ambiguous and patient selection seems to be a crucial step for this device. Further trials are required to estimate the clinical value of this therapeutic concept.

Для цитирования: Andreas Martin, Doll Nicolas, Livesey Steve, Castella Manuel, Kocher Alfred, Casselman Filip, Voth Vladimir, Bannister Christina, Encalada Palacios Juan F, Pereda Daniel, Laufer Guenther, Czesla Markus. Safety and feasibility of a novel adjustable mitral annuloplasty ring: a multicentre European experience.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2015. DOI: 10.1093/ejcts/ezv015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.