Recurrent pericardial effusion caused by pacemaker lead perforation and warfarin therapy at seven years after implantation.
Клинический случайСтатьяПеревод пока не подготовлен
Аннотация
A 66-year-old man was implanted with a pacemaker. Seven years after implantation he was admitted due to cardiogenic cerebral embolism and warfarin therapy was introduced. After that, he suffered recurrent pericardial effusion for unexplained reasons. An exploratory thoracotomy revealed that the screw of the atrial lead had penetrated through the right auricular appendage wall.
Для цитирования: Nakanishi Hiroyuki, Kashiwase Kazunori, Nishio Mayu, Wada Mitsuru, Hirata Akio, Ueda Yasunori. Recurrent pericardial effusion caused by pacemaker lead perforation and warfarin therapy at seven years after implantation.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2012. DOI: 10.1093/europace/eur268.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.