MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/coa.14049

Медицинская библиотека · Clinical otolaryngology : official journal of ENT-UK ; official journal of Netherlands Society for Oto-Rhino-Laryngology & Cervico-Facial Surgery · 2023

Digital image enhancement may improve sensitivity of cholesteatoma detection during endoscopic ear surgery.

Ragonesi Talisa, Niederhauser Laura, Fernandez Ignacio Javier, Molinari Giulia, Caversaccio Marco, Presutti Livio, Anschuetz Lukas

СтатьяРандомизированное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study investigates the possible benefits and limitations of the digital image enhancement systems provided by Storz Professional Image Enhancement System (SPIES) during endoscopic ear surgery (EES) for cholesteatoma. An increased detection of cholesteatoma residuals during the final steps of endoscopic surgery using DIE technology was hypothesized.

DESIGN

Cross-sectional study.

SETTING

Tertiary referral hospital.

METHODS

A total of 10 questionnaires of 18 intraoperative pictures with equal numbers of cholesteatoma and non-cholesteatoma images, each presented in three different image-enhancing modalities (Clara, Spectra A, Spectra B), were generated. Fifty-one experienced ear surgeons participated to the survey and were randomly assigned to a questionnaire and completed it at two time points. The experts were asked to rate for each picture whether cholesteatoma was present or not. The answers were compared with the histopathological reports.

RESULTS

Clara showed the highest accuracy in cholesteatoma detection, followed by Spectra A and lastly Spectra B. In contrast, Spectra B showed the highest sensitivity and Clara the highest specificity, while Spectra A was placed in the middle for both values. Using the Spectra B modality, most responses agreed across the two time points. Ear surgeons assessed the usefulness, as well as preference among image modalities for cholesteatoma surgery, in the following order: Clara, Spectra B, Spectra A.

CONCLUSION

Digital enhancement technologies are applicable to EES. After complete cholesteatoma removal, Spectra B showed the highest sensitivity in the detection of cholesteatoma residuals as compared with Clara and Spectra A. Thus, Spectra B may be recommended to avoid missing any cholesteatoma residuals during EES.

Для цитирования: Ragonesi Talisa, Niederhauser Laura, Fernandez Ignacio Javier, Molinari Giulia, Caversaccio Marco, Presutti Livio, Anschuetz Lukas. Digital image enhancement may improve sensitivity of cholesteatoma detection during endoscopic ear surgery.. Clinical otolaryngology : official journal of ENT-UK ; official journal of Netherlands Society for Oto-Rhino-Laryngology & Cervico-Facial Surgery. 2023. DOI: 10.1111/coa.14049.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.