Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2021

Recapture failure in transcatheter aortic valve replacement with CoreValve Evolut R.

Ito Nobuyasu, Zen Kan, Kuwabara Kensuke, Matoba Satoaki

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Self-expanding prostheses for transcatheter aortic valve replacement (TAVR), which can be recaptured, provide us the option of repositioning for a more accurate placement. We report a very rare case in which the recapture of CoreValve Evolut R (Medtronic, Minneapolis, Minnesota) to correct the implantation depth during the deployment could not be achieved. We planned TAVR with a 23 mm Evolut R prosthesis for a 92-year-old female with severe aortic stenosis and tightly bent thoracic aorta. During the first deployment attempt, the implantation depth was greater than we expected at 2/3 deployment. They tried to recapture and reposition the prosthesis, but the prosthesis was not re-sheathed into the capsule of the delivery system. The prosthesis could not be recaptured despite a repeat attempt, and they were forced to deploy the device as it was. The prosthesis was deployed very carefully and implanted successfully without a pop-up into the ascending aorta. At a later date, this situation was replicated in vitro and was found that the distal segment of the capsule became deformed, increasing the resistance to rotating the grip handle.

Для цитирования: Ito Nobuyasu, Zen Kan, Kuwabara Kensuke, Matoba Satoaki. Recapture failure in transcatheter aortic valve replacement with CoreValve Evolut R.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2021. DOI: 10.1002/ccd.29580.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.