Computed tomography analysis of coronary ostia location following valve-in-valve transcatheter aortic valve replacement with the ACURATE neo valve: Implications for coronary access.

Khokhar Arif A, Laricchia Alessandra, Ponticelli Francesco, Kim Won-Keun, Gallo Francesco, Regazzoli Damiano, Toselli Marco, Sticchi Alessandro, Ruggiero Rossella, Cereda Alberto, Zlahoda-Huzior Adriana, Fisicaro Andrea, Gardi Ilja, Mangieri Antonio, Reimers Bernhard, Dudek Dariusz, Colombo Antonio, Giannini Francesco

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Valve-in-valve transcatheter aortic valve replacement (ViV-TAVR) is an emerging alternative to re-do surgery. However, the challenge of coronary access (CA) following ViV-TAVR is a potential limitation as TAVR expands to younger lower-risk populations.

OBJECTIVES

Using post-implantation computed tomography (CT) scans to evaluate the geometrical relationship between coronary ostia and valve frame in patients undergoing ViV-TAVR with the ACURATE neo valve.

METHODS

Post-implant CT scans of 18 out of 20 consecutive patients treated with the ACURATE neo valve were analyzed. Coronary ostia location in relation to the highest plane (HP) (highest point of the ACURATE neo or surgical valve) was determined. Ostia located below the highest plan were further subclassified according to the gap available between the transcatheter heart valve frame and ostium (transcatheter-to-coronary [TTC] distance). The impact implantation depth has on these geometrical relationships was evaluated.

RESULTS

A total of 21 out of 36 coronary ostia (58%) were located below the level of the HP with the left coronary artery (36%) more likely to be affected than the right (22%). Further sub-classification of these ostia revealed a large (>6 mm), moderate (4-6 mm), and small (<4 mm) TTC distance in 57% (12/21), 38% (8/21), and in 6% (1/18) of cases, respectively. At an implantation depth <4 mm compared to >4 mm, all ostia were located below the HP with no difference in post-procedural mean gradients (14.5 mmHg ± 4.7 vs. 12.6 mmHg ± 5.8; p = .5, 95%CI 3.8-7.5).

CONCLUSIONS

CA following ACURATE neo implantation for ViV-TAVR could potentially be challenging in a significant proportion of patients and specific consideration should be given to the implantation depth.

Для цитирования: Khokhar Arif A, Laricchia Alessandra, Ponticelli Francesco, Kim Won-Keun, Gallo Francesco, Regazzoli Damiano, Toselli Marco, Sticchi Alessandro, Ruggiero Rossella, Cereda Alberto, Zlahoda-Huzior Adriana, Fisicaro Andrea, Gardi Ilja, Mangieri Antonio, Reimers Bernhard, Dudek Dariusz, Colombo Antonio, Giannini Francesco. Computed tomography analysis of coronary ostia location following valve-in-valve transcatheter aortic valve replacement with the ACURATE neo valve: Implications for coronary access.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2021. DOI: 10.1002/ccd.29503.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.