Multi-Axes Lead With Tetrahedral Electrode Tip for Cardiac-Implantable Devices: Creative Concept for Pacing and Sensing Technology.
СтатьяПоддержка исследования: вне правительства СШАОбзорНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Bhaskaran Abhishek, Liang Timothy, Niri Ahmed, Azam Mohammed Ali, Massé Stéphane, Asta John, Magtibay Karl, Lai Patrick F H, Labos Christopher, Ha Andrew C T, Nanthakumar Kumaraswamy
Аннотация
BACKGROUND: We developed a multi-axes lead (MaxLead) incorporating 4 electrodes arranged at the lead-tip, organized in an equidistant tetrahedron. Here, we studied MaxLead performance in sensing, pacing, and activation wavefront-direction analysis.
METHODS: Sixteen explanted animal hearts (from 7 pigs, 7 sheep, and 2 rabbits) were used. Pacing threshold was tested from all axes of MaxLead from right-ventricular (RV) apex before and after simulated dislodgement. In addition, conduction-system pacing was performed in sheep heart preparations from all axes of MaxLead. Sensing via MaxLead positioned at RV apex was tested during sinus rhythm (SR), pacing from RV and left-ventricular (LV) free-wall, and ventricular fibrillation (VF). MaxLead-enabled voltage (MaxV), defined as the largest span of the sensed electric field loop, was compared with traditional lead-tip voltage detection.
RESULTS: Pacing: MaxLead minimized change in pacing threshold owing to lead dislodgement (average voltage change 0.2 mV; 95% confidence interval [CI], -0.5 to 0.9), using multiple bipoles available for pacing. In animals with high conduction system-pacing thresholds (> 2 mV) in 1 or more bipoles (3 of 7), acceptable thresholds (< 1 mV) were demonstrated in an average of 2.5 remaining bipoles. Sensing: MaxV of SR and VF was consistently higher than the highest bipolar voltage (voltage difference averaged -0.18 mV, 95% CI, -0.28 to -0.07), P = 0.001). Electric field-loop geometry consistently differentiated ventricular activation in SR from that during pacing from RV and LV free walls.
CONCLUSIONS: The multi-axes MaxLead electrode showed advantages in pacing, sensing, and mapping and has the potential to allow for improvements in lead-electrode technology for cardiac-implanted electronic devices.
METHODS: Sixteen explanted animal hearts (from 7 pigs, 7 sheep, and 2 rabbits) were used. Pacing threshold was tested from all axes of MaxLead from right-ventricular (RV) apex before and after simulated dislodgement. In addition, conduction-system pacing was performed in sheep heart preparations from all axes of MaxLead. Sensing via MaxLead positioned at RV apex was tested during sinus rhythm (SR), pacing from RV and left-ventricular (LV) free-wall, and ventricular fibrillation (VF). MaxLead-enabled voltage (MaxV), defined as the largest span of the sensed electric field loop, was compared with traditional lead-tip voltage detection.
RESULTS: Pacing: MaxLead minimized change in pacing threshold owing to lead dislodgement (average voltage change 0.2 mV; 95% confidence interval [CI], -0.5 to 0.9), using multiple bipoles available for pacing. In animals with high conduction system-pacing thresholds (> 2 mV) in 1 or more bipoles (3 of 7), acceptable thresholds (< 1 mV) were demonstrated in an average of 2.5 remaining bipoles. Sensing: MaxV of SR and VF was consistently higher than the highest bipolar voltage (voltage difference averaged -0.18 mV, 95% CI, -0.28 to -0.07), P = 0.001). Electric field-loop geometry consistently differentiated ventricular activation in SR from that during pacing from RV and LV free walls.
CONCLUSIONS: The multi-axes MaxLead electrode showed advantages in pacing, sensing, and mapping and has the potential to allow for improvements in lead-electrode technology for cardiac-implanted electronic devices.