DOI 10.1016/j.cjca.2021.07.013
Multi-Axes Lead With Tetrahedral Electrode Tip for Cardiac-Implantable Devices: Creative Concept for Pacing and Sensing Technology.
Аннотация
BACKGROUND
We developed a multi-axes lead (MaxLead) incorporating 4 electrodes arranged at the lead-tip, organized in an equidistant tetrahedron. Here, we studied MaxLead performance in sensing, pacing, and activation wavefront-direction analysis.
METHODS
Sixteen explanted animal hearts (from 7 pigs, 7 sheep, and 2 rabbits) were used. Pacing threshold was tested from all axes of MaxLead from right-ventricular (RV) apex before and after simulated dislodgement. In addition, conduction-system pacing was performed in sheep heart preparations from all axes of MaxLead. Sensing via MaxLead positioned at RV apex was tested during sinus rhythm (SR), pacing from RV and left-ventricular (LV) free-wall, and ventricular fibrillation (VF). MaxLead-enabled voltage (MaxV), defined as the largest span of the sensed electric field loop, was compared with traditional lead-tip voltage detection.
RESULTS
Pacing: MaxLead minimized change in pacing threshold owing to lead dislodgement (average voltage change 0.2 mV; 95% confidence interval [CI], -0.5 to 0.9), using multiple bipoles available for pacing. In animals with high conduction system-pacing thresholds (> 2 mV) in 1 or more bipoles (3 of 7), acceptable thresholds (< 1 mV) were demonstrated in an average of 2.5 remaining bipoles. Sensing: MaxV of SR and VF was consistently higher than the highest bipolar voltage (voltage difference averaged -0.18 mV, 95% CI, -0.28 to -0.07), P = 0.001). Electric field-loop geometry consistently differentiated ventricular activation in SR from that during pacing from RV and LV free walls.
CONCLUSIONS
The multi-axes MaxLead electrode showed advantages in pacing, sensing, and mapping and has the potential to allow for improvements in lead-electrode technology for cardiac-implanted electronic devices.