QT Interval Monitoring with Handheld Heart Rhythm ECG Device in COVID-19 Patients.

Minguito-Carazo Carlos, Echarte-Morales Julio, Benito-González Tomás, Del Castillo-García Samuel, Rodríguez-Santamarta Miguel, Sánchez-Muñoz Enrique, Maniega Clea González, García-Bergel Rubén, Menéndez-Suárez Paula, Prieto-González Silvia, Palacios-Echavarren Carmen, Borrego-Rodríguez Javier, Flores-Vergara Guisela, Iglesias-Garriz Ignacio, Fernández-Vázquez Felipe

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

QTc prolongation is an adverse effect of COVID-19 therapies. The use of a handheld device in this scenario has not been addressed.

OBJECTIVES

To evaluate the feasibility of QTc monitoring with a smart device in COVID-19 patients receiving QTc-interfering therapies.

METHODS

Prospective study of consecutive COVID-19 patients treated with hydroxychloroquine ± azithromycin ± lopinavir-ritonavir. ECG monitoring was performed with 12-lead ECG or with KardiaMobile-6L. Both registries were also sequentially obtained in a cohort of healthy patients. We evaluated differences in QTc in COVID-19 patients between three different monitoring strategies: 12-lead ECG at baseline and follow-up (A), 12-lead ECG at baseline and follow-up with the smart device (B), and fully monitored with handheld 6-lead ECG (group C). Time needed to obtain an ECG registry was also documented.

RESULTS

One hundred and eighty-two COVID-19 patients were included (A: 119(65.4%); B: 50(27.5%); C: 13(7.1%). QTc peak during hospitalization did significantly increase in all groups. No differences were observed between the three monitoring strategies in QTc prolongation (p = 0.864). In the control group, all but one ECG registry with the smart device allowed QTc measurement and mean QTc did not differ between both techniques (p = 0.612), displaying a moderate reliability (ICC 0.56 [0.19-0.76]). Time of ECG registry was significantly longer for the 12-lead ECG than for handheld device in both cohorts (p < 0.001).

CONCLUSION

QTc monitoring with KardiaMobile-6L in COVID-19 patients was feasible. Time of ECG registration was significantly lower with the smart device, which may offer an important advantage for prevention of virus dissemination among healthcare providers.

Для цитирования: Minguito-Carazo Carlos, Echarte-Morales Julio, Benito-González Tomás, Del Castillo-García Samuel, Rodríguez-Santamarta Miguel, Sánchez-Muñoz Enrique, Maniega Clea González, García-Bergel Rubén, Menéndez-Suárez Paula, Prieto-González Silvia, Palacios-Echavarren Carmen, Borrego-Rodríguez Javier, Flores-Vergara Guisela, Iglesias-Garriz Ignacio, Fernández-Vázquez Felipe. QT Interval Monitoring with Handheld Heart Rhythm ECG Device in COVID-19 Patients.. Global heart. 2021. DOI: 10.5334/gh.916.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.