m+Medelevator.RU
← Поиск публикаций
Magnetic resonance in medicine · 2015

Ensuring safety and functionality of electroglottography measurements during dynamic pulmonary MRI.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Özen Ali Caglar, Traser Louisa, Echternach Matthias, Dadakova Tetiana, Burdumy Michael, Richter Bernhard, Bock Michael

Аннотация

PURPOSE: To combine vocal tract measurements with dynamic MRI of the lungs to provide fundamental insights into the lung physiology during singing.

METHODS: To analyze vocal fold oscillatory patterns during dynamic lung MRI, an electroglottography (EGG) system was modified to allow for simultaneous EGG measurements during MR image acquisitions. A low-pass filter was introduced to suppress residual radiofrequency (RF) coupling into the EGG signal. RF heating was tested in a gel phantom to ensure MR safety, and functionality of the EGG device was assessed in a volunteer experiment at singing frequencies from A5 to A3. In the recorded EGG signals, remaining RF interferences were removed by independent component analysis post processing, and standard EGG parameters such as fundamental frequency, contact quotient and jitter were calculated. In a second volunteer experiment, EGG recordings were compared with lung diameter measurements from 2D time-resolved trueFISP acquisitions.

RESULTS: RF heating measurements resulted in less than 1.2°C temperature increase in the gel phantom. EGG parameters measured during MR imaging are within the range of ideal values. In the lung measurement, both the lung diameter and the EGG recordings could be successfully performed with only minimal interference.

CONCLUSION: EGG recording is pos sible during dynamic lung MRI, and glottal activity can be studied safely at 1.5T. Magn Reson Med 76:1629-1635, 2016. © 2015 International Society for Magnetic Resonance in Medicine.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты