BACKGROUND
Recent guidelines for infant cardiopulmonary resuscitation emphasize that all rescuers should minimize interruption of chest compressions, even for endotracheal intubation. We compared the utility of the Pentax-AWS Airwayscope (AWS) with the Glidescope (GS) during chest compressions on an infant manikin.
METHODS
Twenty-four anesthesiologists with more than two years of experience performed tracheal intubation on an infant manikin using the AWS and GS, with or without chest compressions.
RESULTS
In GS trials, none of the participants failed without compressions, while three failed with compressions. In AWS trials, all participants succeeded regardless of chest compressions. Intubation time was significantly longer with chest compressions with the GS (P < 0.05), but not with the AWS. Difficulty of operation on a visual analog scale (VAS) for laryngoscopy did not increase significantly with chest compressions with either the GS or the AWS, while the VAS for tube passage through the glottis increased with compressions with the GS, but not with the AWS.
CONCLUSION
We conclude that in infant simulations managed by anesthesiologists, the AWS performed better than the GS for endotracheal intubation with chest compressions.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.