DOI 10.1097/00004728-200101000-00026
Carpal tunnel syndrome: assessment by turbo spin echo, spin echo, and magnetization transfer imaging applied in a low-field MR system.
Аннотация
PURPOSE: The purpose of this work was to evaluate patients with carpal tunnel syndrome (CTS) using a low-field extremity MR system (E-MRI: 0.2 T).
METHOD: Twenty-two patients with typical findings of CTS and 30 control persons were imaged on an E-MRI. Axial T2-weighted turbo SE (TSE), T1-weighted SE sequences, and 2D GRE magnetization transfer (MTC) sequences were compared. SE and MTC sequences were obtained before and after contrast agent administration (0.1 mmol/kg body wt of Gd-DTPA). Two readers evaluated typical MR findings of CTS independently.
RESULTS
Patients with CTS demonstrated palmar bowing of the flexor retinaculum significantly more often. The normal or edematous median nerve was best identified on TSE and MTC scans (kappa = 0.59 and 0.8). The MTC sequences showed perineural enhancement significantly better than respective T1-weighted SE sequences but were rated second in comparison with T2-weighted TSE scans.
CONCLUSION
At low-field strength, median nerve edema is best depicted on T2-weighted TSE sequences, whereas MTC sequences are most sensitive to perineural contrast enhancement.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.