ON-X bileaflet valve in aortic position--early experience shows an improved hemodynamic profile.
Аннотация
BACKGROUND
Despite considerable progress during the last decades, mechanical heart valves still have significant disadvantages in performance compared to native valves. To optimize the hemodynamic profile, a new mechanical bileaflet valve (ON-X) was developed and introduced in 1997. It was the aim of this study to assess the feasibility, safety, and the early postoperative hemodynamics with this new valve in vivo.
METHODS
We analyzed 19 patients (11 male, 8 female; 63.2 +/- 8.2 years; aortic stenosis: n = 13; aortic incompetence: n = 2; combined aortic lesion: n = 2), undergoing aortic valve replacement with this prosthesis (valve size 19 mm: n = 1; 21 mm: n = 6; 23 mm: n = 8; 25 mm: n = 4; additional CABG: n = 4). In addition to intraoperative pressure gradient measurements, echocardiography was performed early postoperatively and after 3 months, evaluating pressure loss, effective orifice area, and regression of left-ventricular hypertrophy.
RESULTS
No major perioperative complications were observed. The echocardiographic evaluation demonstrated a significantly increased effective orifice area and lower transvalvular gradients in all valve sizes compared with literature values for the St. Jude Medical prosthesis. Effective regression of left-ventricular hypertrophy (-23.6%) was observed over the first three months.
CONCLUSIONS
These preliminary data confirm the favorable hemodynamic characteristics of the ON-X valve as compared to standard bileaflet designs. Implantation was feasible and safe, and no major postoperative complications such as thrombembolic events were observed.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.