Subcutaneous Implantable Cardioverter Defibrillator Lead Repositioning for Preventing Inappropriate Shocks Due to Myopotential Oversensing in a Post-Fulminant Myocarditis Patient.

Sasaki Takehito, Nakamura Kohki, Naito Shigeto

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 28-year-old female presented with fulminant lymphocytic myocarditis. She developed cardiogenic shock, frequent sustained ventricular tachycardia, and fibrillation (VT and VF). The left ventricular ejection fraction improved from 5% to 40% after medical therapy, but the right ventricular systolic dysfunction and enlargement persisted. In addition, sustained VTs, requiring direct current cardioversion, occurred during oral administration of amiodarone following intravenous amiodarone, even after percutaneous stellate ganglion block. Standard body surface electrocardiogram (ECG) screening for an implantation of a subcutaneous implantable cardioverter-defibrillator (S-ICD) (EMBLEM™ S-ICD, Boston Scientific, Marlborough, MA, USA) demonstrated that two of the three sensing vectors were eligible in spite of very low-amplitude QRS complexes in the body surface ECGs. After implantation of the S-ICD, the patient experienced repetitive, inappropriate shocks due to pectoral myopotential oversensing, which could not be resolved by reprogramming the device settings. Thus, the S-ICD lead was changed from the standard left parasternal position to the midline of the sternum to reduce muscular noise due to myopotentials. Thereafter, the patient experienced appropriate ICD shocks for sustained VT and VF but no inappropriate ICD sensing or shocks. Lead repositioning may be one of the feasible solutions in S-ICD patients with low-amplitude QRS complexes and inappropriate shocks due to myopotential oversensing which cannot be resolved by reprogramming the device settings.

Для цитирования: Sasaki Takehito, Nakamura Kohki, Naito Shigeto. Subcutaneous Implantable Cardioverter Defibrillator Lead Repositioning for Preventing Inappropriate Shocks Due to Myopotential Oversensing in a Post-Fulminant Myocarditis Patient.. International heart journal. 2019. DOI: 10.1536/ihj.18-324.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.