DOI 10.1016/j.jcin.2021.08.058
Vascular Quality Initiative Surveillance of Femoropopliteal Artery Paclitaxel Devices.
Аннотация
OBJECTIVES
This study sought to determine the safety and efficacy of paclitaxel (PTX) devices in the treatment of peripheral artery disease involving the femoropopliteal artery.
BACKGROUND
A meta-analysis of PTX devices for the treatment of femoropopliteal artery disease reported a mortality signal.
METHODS
This was a multicenter cohort study using an integrated clinical data surveillance system to conduct a prospective, propensity score-matched survival analysis of 2,456 patients in the Society for Vascular Surgery Vascular Quality Initiative from January 2017 to May 2020. The study compared PTX drug-coated balloon angioplasty versus percutaneous transluminal balloon angioplasty, PTX drug-eluting stents versus bare-metal stents, and any PTX device versus any non-PTX device. The primary outcome was 2-year survival. Secondary endpoints were successful ambulation and interventional success.
RESULTS
Treatment with any PTX device versus any non-PTX device was associated with increased 2-year survival (89.5% vs 86.7%; HR: 0.79; 95% CI: 0.72-0.87; P = 0.004), improved interventional success (81.6% vs 77.6%; HR: 0.82; 95% CI: 0.74-0.91; P < 0.001), and higher rates of independent ambulation at 1 year (86.0% vs 83.4%; HR: 0.85; 95% CI: 0.79-0.91; P = 0.008). Treatment with PTX drug-coated balloon angioplasty was associated with improved survival at 2 years (88.9% vs 85.7%; HR: 0.77; 95% CI: 0.70-0.86; P = 0.005), while PTX drug-eluting stent therapy was associated with similar survival compared with bare-metal stent therapy (91.3% vs 89.6%; HR: 0.84; 95% CI: 0.70-1.01; P = 0.36).
CONCLUSIONS
In this prospective, active surveillance of a national clinical registry, PTX-containing devices were associated with increased survival at 2 years and improved clinical outcomes at 1 year. (VQI DELTA Paclitaxel Device Safety Analysis [VQI-PTX]; NCT04110288).
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.