Refined heart failure detection algorithm for improved clinical reliability of OptiVol alerts in CRT-D recipients.

Vamos Mate, Nyolczas Noemi, Bari Zsolt, Bogyi Peter, Muk Balazs, Szabo Barna, Ancsin Bettina, Kiss Robert G, Duray Gabor Z

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The reliability of intrathoracic impedance monitoring for prediction of heart failure (HF) by implantable cardiac devices is controversial. Despite using additional device-based parameters described in the PARTNERS HF study, such as new onset of arrhythmias, abnormal autonomics, low biventricular pacing rate or patient activity level, the predictive power of device diagnostic algorithm is still in doubt. The objective of this study was to compare the device diagnostic algorithm described in the PARTNERS HF study to a newly developed algorithm applying refined diagnostic criteria.

METHODS

Fourty two patients were prospectively enrolled who had been implanted with an intrathoracic impedance and remote monitoring capable implantable cardiac defibrillator with a cardiac resychroniza-tion therapy (CRT-D) device in this observational study. If a remote OptiVolTM alert occurred, patients were checked for presence of HF symptoms. A new algorithm was derived from the original PARTNERS HF criteria, considering more sensitive cut-offs and changes of patterns of the device-based parameters.

RESULTS

During an average follow-up of 38 months, 722 remote transmissions were received. From the total of 128 transmissions with OptiVol alerts, 32 (25%) corresponded to true HF events. Upon multivariate discriminant analysis, low patient activity, high nocturnal heart rate, and low CRT pacing (< 90%) proved to be independent predictors of true HF events (all p < 0.01). Incorporating these three refined criteria in a new algorithm, the diagnostic yield of OptiVol was improved by increasing specific-ity from 37.5% to 86.5%, positive predictive value from 34.1% to 69.8% and area under the curve from 0.787 to 0.922 (p < 0.01), without a relevant loss in sensitivity (96.9% vs. 93.8%).

CONCLUSIONS

A refined device diagnostic algorithm based on the parameters of low activity level, high nocturnal heart rate, and suboptimal biventricular pacing might improve the clinical reliability of OptiVol alerts.

Для цитирования: Vamos Mate, Nyolczas Noemi, Bari Zsolt, Bogyi Peter, Muk Balazs, Szabo Barna, Ancsin Bettina, Kiss Robert G, Duray Gabor Z. Refined heart failure detection algorithm for improved clinical reliability of OptiVol alerts in CRT-D recipients.. Cardiology journal. 2017. DOI: 10.5603/cj.a2017.0077.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.