High-Intensity Focused Ultrasound Therapy in the Uterine Fibroid: A Clinical Case Study of Poor Heating Efficacy.

Suomi Visa, Viitala Antti, Sainio Teija, Komar Gaber, Sequeiros Roberto Blanco

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

A clinical case study of high-intensity focused ultrasound (HIFU) treatment in the uterine fibroid was conducted. During the therapy, poor heating efficacy was observed which could be attributed to several factors such as the local perfusion rate, patient-specific anatomy or changes in acoustic parameters of the ultrasound field. In order to determine the cause of the diminished heating, perfusion analyses and ultrasound simulations were conducted using the magnetic resonance imaging (MRI) data from the treatment. The perfusion analysis showed high local perfusion rate in the myoma (301.0 ± 25.6 mL/100 g/min) compared to the surrounding myometrium (233.8 ± 16.2 mL/100 g/min). The ultrasound simulations did not show large differences in the focal point shape or the acoustic pressure (2.07 ± 0.06 MPa) when tilting the transducer. However, a small shift (-2.2 ± 1.3 mm) in the axial location of the focal point was observed. The main causes for the diminished heating were likely the high local perfusion and ultrasound attenuation due to the deep location of the myoma.

Для цитирования: Suomi Visa, Viitala Antti, Sainio Teija, Komar Gaber, Sequeiros Roberto Blanco. High-Intensity Focused Ultrasound Therapy in the Uterine Fibroid: A Clinical Case Study of Poor Heating Efficacy.. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2019. DOI: 10.1109/embc.2019.8857147.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.