DOI 10.1016/j.echo.2011.04.009

Медицинская библиотека · Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography · 2011

Intracardiac echocardiography: a new guiding tool for transcatheter aortic valve replacement.

Bartel Thomas, Bonaros Nikolaos, Müller Ludwig, Friedrich Guy, Grimm Michael, Velik-Salchner Corinna, Feuchtner Gudrun, Pedross Florian, Müller Silvana

Сравнительное исследованиеСтатьяРандомизированное исследованиеVideo-Audio MediaПеревод пока не подготовлен

Аннотация

BACKGROUND

Echocardiography has been debated as an adjunct for transcatheter aortic valve replacement (TAVR). The aim of this prospective study was to comparatively evaluate intraprocedural guidance using intracardiac echocardiography (ICE) and transesophageal echocardiography (TEE).

METHODS

Fifty high-risk patients with severe aortic stenosis scheduled for TAVR were randomized to either guidance using ICE (group 1; n = 25) or monitoring using TEE (group 2; n = 25).

RESULTS

In contrast to TEE, ICE allowed continuous monitoring. The need for probe repositioning during the procedure was much lower in group 1 (0.1 ± 0.3 vs 5.7 ± 0.7 maneuvers, P < .001). Compared with TEE, the transcaval intracardiac echocardiographic view provided higher coaxiality with the ascending aorta expressed as the length of ascending aorta depicted (4.9 ± 1.2 vs 6.1 ± 1.2 cm, P = .003). Both coronary ostia were more frequently visualized in group 1 (18 vs 2 cases, P < .001). ICE-derived annular measurements were correlated closely with preinterventional readings on TEE (n = 25, r(2) = 0.90, P < .001). TEE underestimated intraprocedural pressure gradients in comparison with preinterventional measurements (mean difference, -10.2 ± 11.1 mm Hg; n = 11, P = .012), but ICE did not (mean difference, -0.3 ± 14.1 mm Hg; n = 25, P = .913). ICE and TEE detected newly grown thrombi (2 vs 1 case). Severe complications (e.g., annular dissection, pericardial effusion) were not observed.

CONCLUSIONS

ICE, which is compatible with sedation and local anesthesia, can be considered an alternative to TEE for intraprocedural guidance during TAVR. It also seems to match the required work flow during TAVR better than TEE.

Для цитирования: Bartel Thomas, Bonaros Nikolaos, Müller Ludwig, Friedrich Guy, Grimm Michael, Velik-Salchner Corinna, Feuchtner Gudrun, Pedross Florian, Müller Silvana. Intracardiac echocardiography: a new guiding tool for transcatheter aortic valve replacement.. Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography. 2011. DOI: 10.1016/j.echo.2011.04.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.