DOI 10.1046/j.1540-8167.2002.00990.x
Comparison of electrocardiogram and intrathoracic electrogram signals for detection of ischemic ST segment changes during normal sinus and ventricular paced rhythms.
Аннотация
INTRODUCTION
The aim of this study was to compare surface ECGs with electrograms (EGM) that are available from implanted devices for the ability to detect ischemic ST segment changes during normal sinus (NS) and ventricular paced (VP) rhythms.
METHODS AND
RESULTS
ECG leads I, II, and V2, right atrial ring to left pectoral patch (representing the can of the device), right ventricular ring to left pectoral patch, and right atrial ring to right ventricular ring EGM were recorded continuously during percutaneous transluminal coronary angioplasty. One balloon inflation (> or = 60 sec) was analyzed from each of 22 NS and 22 VP subjects. The parameter AST was defined as the maximum absolute ST segment deviation (from isoelectric) during the first 60 seconds of inflation, measured relative to the baseline (preinflation) ST segment deviation. For EGM, a normalized deltaST was defined as the AST divided by the ratio of QRS amplitudes of EGM to ECG. During NS, the deltaST for EGM (0.43 mV) was significantly larger than that of ECG (0.09 mV, P = 0.0001) but the normalized deltaST for EGM (0.11 mV) was comparable to that of ECG (0.09 mV, P = 0.45). During VP, the AST for EGM (1.08 mV) was significantly larger than that of ECG (0.17 mV, P = 0.0001), but the normalized AST for EGM (0.11 mV) was significantly smaller than that of ECG (0.17 mV, P = 0.02).
CONCLUSION
During both NS and VP, ischemic ST segment changes were significantly larger in EGM than in ECG. Much of this difference appears to be related to larger amplitudes of EGM signals. (J
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.