Kuramitsu Shoichi, Iwabuchi Masashi, Haraguchi Takuya, Domei Takenori, Nagae Ayumu, Hyodo Makoto, Yamaji Kyohei, Soga Yoshimitsu, Arita Takeshi, Shirai Shinichi, Kondo Katsuhiro, Ando Kenji, Sakai Koyu, Goya Masahiko, Takabatake Yoshitaka, Sonoda Shinjo, Yokoi Hiroyoshi, Toyota Fumitoshi, Nosaka Hideyuki, Nobuyoshi Masakiyo

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Stent fracture (SF) after drug-eluting stent implantation has recently become an important concern because of its potential association with in-stent restenosis and stent thrombosis. However, the incidence and clinical impact of SF after everolimus-eluting stent implantation remain unclear.

METHODS AND

RESULTS

A total of 1035 patients with 1339 lesions undergoing everolimus-eluting stent implantation and follow-up angiography 6 to 9 months after index procedure were analyzed. SF was defined as complete or partial separation of the stent, as assessed by plain fluoroscopy or intravascular ultrasound during follow-up. We assessed the rates of SF and major adverse cardiac events, defined as cardiac death, myocardial infarction, stent thrombosis, and clinically driven target lesion revascularization within 9 months. SF was observed in 39 of 1339 lesions (2.9%) and in 39 of 1035 patients (3.8%). Ostial stent location and lesions with hinge motion, tortuosity, or calcification were independent predictors of SF. The rate of myocardial infarction and target lesion revascularization were significantly higher in the SF group than in the non-SF group (5.1% versus 0.4%; P=0.018 and 25.6% versus 2.0%; P<0.001, respectively). Stent thrombosis was more frequently observed in the SF group than in the non-SF group (5.1% versus 0.4%; P=0.018). Major adverse cardiac events within 9 months were significantly higher in the SF group than in the non-SF group (25.6% versus 2.3%; P<0.001).

CONCLUSIONS

SF after everolimus-eluting stent implantation occurs in 2.9% of lesions and is associated with higher rate of major adverse cardiac events, driven by higher target lesion revascularization and stent thrombosis.

Для цитирования: Kuramitsu Shoichi, Iwabuchi Masashi, Haraguchi Takuya, Domei Takenori, Nagae Ayumu, Hyodo Makoto, Yamaji Kyohei, Soga Yoshimitsu, Arita Takeshi, Shirai Shinichi, Kondo Katsuhiro, Ando Kenji, Sakai Koyu, Goya Masahiko, Takabatake Yoshitaka, Sonoda Shinjo, Yokoi Hiroyoshi, Toyota Fumitoshi, Nosaka Hideyuki, Nobuyoshi Masakiyo. Incidence and clinical impact of stent fracture after everolimus-eluting stent implantation.. Circulation. Cardiovascular interventions. 2012. DOI: 10.1161/circinterventions.112.969238.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.