MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2012.07.002

Медицинская библиотека · Heart rhythm · 2012

Implantable cardioverter-defibrillator FDA safety advisories: Impact on patient mortality and morbidity.

Sengupta Jay, Kendig Arthur C, Goormastic Marlene, Hwang Eui-Seock, Ching Elizabeth A, Chung Roy, Lindsay Bruce D, Tchou Patrick J, Wilkoff Bruce L, Niebauer Mark J, Martin David O, Varma Niraj, Wazni Oussama, Saliba Walid, Kanj Mohamed, Bhargava Mandeep, Dresing Thomas, Taigen Tyler, Ingelmo Christopher, Bassiouny Mohamed, Cronin Edmond M, Wilsmore Bradley, Rickard John, Chung Mina K

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

A significant proportion of implantable cardioverter-defibrillators (ICDs) have been subject to Food and Drug Administration (FDA) advisories. The impact of device advisories on mortality or patient care is poorly understood. Although estimated risks of ICD generators under advisory are low, dependency on ICD therapies to prevent sudden death justifies the assessment of long-term mortality.

OBJECTIVE

To test the association of FDA advisory status with long-term mortality.

METHODS

The study was a retrospective, single-center review of clinical outcomes, including device malfunctions, in patients from implantation to either explant or death. Patients with ICDs first implanted at Cleveland Clinic between August 1996 and May 2004 who became subject to FDA advisories on ICD generators were identified. Mortality was determined by using the Social Security Death Index.

RESULTS

In 1644 consecutive patients receiving first ICD implants, 704 (43%) became subject to an FDA advisory, of which 172 (10.5%) were class I and 532 (32.3%) were class II. ICDs were explanted before advisory notifications in 14.0% of class I and 10.1% of class II advisories. Among ICDs under advisory, 28 (4.0%) advisory-related and 15 non-advisory- related malfunctions were documented. Over a median follow-up of 70 months, 814 patients died. Kaplan-Meier 5-year survival rate was 65.6% overall, and 64.2, 61.1, and 69.3% in patients with no, class I, and class II advisories, respectively (P = .17).

CONCLUSIONS

ICD advisories impacted 43% of the patients. Advisory-related malfunctions affected 4% within the combined advisory group. Based on a conservative management strategy, ICDs under advisory were not associated with increased mortality over a background of significant disease-related mortality.

Для цитирования: Sengupta Jay, Kendig Arthur C, Goormastic Marlene, Hwang Eui-Seock, Ching Elizabeth A, Chung Roy, Lindsay Bruce D, Tchou Patrick J, Wilkoff Bruce L, Niebauer Mark J, Martin David O, Varma Niraj, Wazni Oussama, Saliba Walid, Kanj Mohamed, Bhargava Mandeep, Dresing Thomas, Taigen Tyler, Ingelmo Christopher, Bassiouny Mohamed, Cronin Edmond M, Wilsmore Bradley, Rickard John, Chung Mina K. Implantable cardioverter-defibrillator FDA safety advisories: Impact on patient mortality and morbidity.. Heart rhythm. 2012. DOI: 10.1016/j.hrthm.2012.07.002.