MedElevator Войти в исследования
← Поиск публикаций
Медицинская библиотека · The Journal of invasive cardiology · 2012

Feasibility and safety of the second-generation, frequency domain optical coherence tomography (FD-OCT): a multicenter study.

Yoon Joo Heung, Di Vito Luca, Moses Jeffrey W, Fearon William F, Yeung Alan C, Zhang Shaosong, Bezerra Hiram G, Costa Marco A, Jang Ik-Kyung

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study sought to assess the effectiveness and safety of the second-generation frequency-domain optical coherence tomography (FD-OCT) system.

BACKGROUND

The second-generation FD-OCT was recently developed, with simplified imaging technique and faster acquisition time compared to the first-generation time-domain OCT. However, the safety and effectiveness of the FD-OCT has not been evaluated, and this study was conceived as a preapproval study for Food and Drug Administration clearance for clinical use in the United States.

METHODS

A total of 50 patients were enrolled from 3 institutions. Following stent implantation, the FD-OCT was performed with contrast injection through the guiding catheter to acquire pullback images with the pressure-triggered automatic pullback device. The primary endpoint was to achieve a median clear image length of more than 24 mm. Serious procedure-related or postprocedural adverse events (death, myocardial infarction, or ventricular arrhythmia) were recorded to assess safety of the device.

RESULTS

The primary endpoint of obtaining >24 mm of median clear image length (CIL) was achieved in 94% of the subjects (47 out of 50), with measured CIL of 43.2 mm. In 5 patients (10.6%), a second attempt was necessary due to suboptimal image quality of the first pullback. In 36 patients (76.6%), a full stent length was obtained during the first attempt. There were no serious procedure-related or postprocedural adverse events.

CONCLUSIONS

The new second-generation FD-OCT system provides fast and reliable resolution images of the coronary artery. The pullback can be safely performed over long segments of the artery without serious adverse events.

Для цитирования: Yoon Joo Heung, Di Vito Luca, Moses Jeffrey W, Fearon William F, Yeung Alan C, Zhang Shaosong, Bezerra Hiram G, Costa Marco A, Jang Ik-Kyung. Feasibility and safety of the second-generation, frequency domain optical coherence tomography (FD-OCT): a multicenter study.. The Journal of invasive cardiology. 2012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.